6 Troubleshooting
6 Troubleshooting
6 Troubleshooting
6 Troubleshooting
This table provides procedures for identifying and correcting operator
problems. This list includes only those problems correctable without
special equipment. Refer problems beyond the scope of this section to
an authorized Nellcor Puritan Bennett Service Representative.
Troubleshooting Chart
Troubleshooting Chart
Troubleshooting Chart
Troubleshooting Chart
Symptom
Symptom
Symptom
Symptom
Cause
Cause
Cause
Cause
Solution
Solution
Solution
Solution
Green AC power light is out with
unit connected to AC power
AC Power Switch
Move switch to ON
No AC power at outlet
Check for 110 VAC at outlet
Open Circuit Breaker
Reset by moving AC switch to ON
Defective cord or connector
Check cord and connector for possi-
ble loose connections.
Defective LED
Contact Nellcor Puritan Bennett
Power light ON, but ventilator
fails to cycle
Mode switch in Standby posi-
tion
Move MODE switch to an active
position
Alarms activate when unit is
turned on
Normal test of audible and
visual indicators
Not applicable; Alarms automati-
cally reset
Low Pressure alarm sounds fre-
quently
Incorrect alarm settings
Check alarm control setting
Improper patient connections
Check patient connection (trach)
for a secure, leakproof fit
Leaks in components or con-
nections
Check all connections, especially
the humidifier and exhalation valve
Water in small tubing
Remove water
No patient circuit is con-
nected to unit; no pressure is
registered
Connect circuit with test lung to
achieve proper pressure according
to alarm set point
Pressure Limit control is over-
tightened (LP10, LP20)
Do not overtighten the Pressure Limit
control