Specifications
Specifications
Specifications
Specifications
LP6 Plus, LP10, and LP20 Technical Manual
November 2000
Page 2-13
Page 2-13
Page 2-13
Page 2-13
Pressure Limit
Pressure Limit
Pressure Limit
Pressure Limit (LP10, LP20):
(LP10, LP20):
(LP10, LP20):
(LP10, LP20):
Locking knob sets pressure limit level from 15-50 cmH
2
O or Closes
off the pressure limit feature
Input
Input
Input
Input
Patient Pressure:
Patient Pressure:
Patient Pressure:
Patient Pressure:
Port for connection to the proximal pressure line of the patient cir-
cuit
Outputs
Outputs
Outputs
Outputs
Patient Air
Patient Air
Patient Air
Patient Air::::
22 mm tube for connection to the bacteria filter
Exhalation Valve:
Exhalation Valve:
Exhalation Valve:
Exhalation Valve:
Port for connection to the exhalation valve of the patient circuit
Indicators
Indicators
Indicators
Indicators
Normal Events
Normal Events
Normal Events
Normal Events
Patient Pressure Meter:
Displays patient pressure -10 to +100
cmH
2
O; also displays battery charge and machine hours of opera-
tion when appropriate buttons are pressed.
Breathing Effort:
Green LED activated by adequate patient
breathing effort.
Power:
LEDs to indicate operating power source: green AC Pwr/Batt
Charge, amber External Battery, flashing amber Internal Battery.
Alarms:
Alarms:
Alarms:
Alarms:
Flashing Red LEDs:
Low Pressure/Apnea, Low Power, High Pres-
sure, Setting Error, Power Switchover
Audible Alarms:
Audible Alarms:
Audible Alarms:
Audible Alarms:
Double-Pulsed Tone:
Low Pressure/Apnea (LP20 only)
Pulsating Tone:
Low Pressure/Apnea (LP6 Plus, LP10), Low Power,
High Pressure, Setting Error, Power Switchover
Steady Tone:
Loss of microprocessor control, depleted internal bat-
tery
Reminder Tone:
Every five min. when powered by internal battery,
and each time accessory printer generates a report
Rear Panel Controls
Rear Panel Controls
Rear Panel Controls
Rear Panel Controls
AC Circuit Breaker:
AC Circuit Breaker:
AC Circuit Breaker:
AC Circuit Breaker:
Internal to the AC power switch (1 for on, 0 for off)
Inputs
Inputs
Inputs
Inputs
Inlet Filter:
Inlet Filter:
Inlet Filter:
Inlet Filter:
Intake for patient air, screw-off cap for filter change
External Battery:
External Battery:
External Battery:
External Battery:
Connection for 12 VDC battery
Communication Port
Communication Port
Communication Port
Communication Port
Calibration information during service procedure
Outputs
Outputs
Outputs
Outputs
Vent
Vent
Vent
Vent::::
Cooling vent for internal ventilator components
Communications Port
Communications Port
Communications Port
Communications Port::::
Connection for optional printer or display
Alarm:
Alarm:
Alarm:
Alarm:
Connection for Remote Alarm accessory
Alarm Output: (LP20 only)
Alarm Output: (LP20 only)
Alarm Output: (LP20 only)
Alarm Output: (LP20 only)
Factory default is normally open relay, 1.67 Hz 50% duty cycle
open/closed for Low Pressure alarms; closed for all other alarm con-
ditions and when mode is set to Standby. Can be changed to nor-
mally closed; call Technical Service (800 497-3787) for details. 32
VDC @ 0.5 A