Specifications
Specifications
Specifications
Specifications
LP6 Plus, LP10, and LP20 Technical Manual
November 2000
Page 2-12
Page 2-12
Page 2-12
Page 2-12
Specifications
Specifications
Specifications
Specifications
LP6 Plus, LP10, and LP20 Volume Ventilators
LP6 Plus, LP10, and LP20 Volume Ventilators
LP6 Plus, LP10, and LP20 Volume Ventilators
LP6 Plus, LP10, and LP20 Volume Ventilators
Power Line
Power Line
Power Line
Power Line
110 VAC range (100-127 VAC) or 220 VAC range (220-240 VAC), 50/
60 Hz, external voltage selector switch.
Power Usage
Power Usage
Power Usage
Power Usage
Maximum:
Maximum:
Maximum:
Maximum:
630 kW per year
Nominal:
Nominal:
Nominal:
Nominal:
315 kW per year
Extension Cord Gauge:
Extension Cord Gauge:
Extension Cord Gauge:
Extension Cord Gauge:
Use three-conductor cord only; up to 49’, use 18 gauge cord; up to
99’, use 16 gauge cord; up to 200’, use 14 gauge cord
External Battery:
External Battery:
External Battery:
External Battery:
Approximately 20 hrs operation with 75-80 amp-hour 12 VDC deep-
cycle, gel-cell battery. Approximately 10 hrs operation with 35-40
amp-hour 12 VDC deep-cycle, gel-cell battery
Internal Battery:
Internal Battery:
Internal Battery:
Internal Battery:
Approximately one hour operation
Type
Type
Type
Type
Volume ventilator
Motor:
Motor:
Motor:
Motor:
Brushless induction
Pump:
Pump:
Pump:
Pump:
Piston, 100-2200 ml tidal volume capability
Front Panel Controls
Front Panel Controls
Front Panel Controls
Front Panel Controls
Alarm Silence/Reset:
Alarm Silence/Reset:
Alarm Silence/Reset:
Alarm Silence/Reset:
Push-button to pre-silence alarms before events (LP20 only), silence
alarms during events, or reset alarms after events; used with Bat-
tery Test button to read machine operating hours on Patient Pres-
sure meter
Battery Test:
Battery Test:
Battery Test:
Battery Test:
Push-button to show battery charge level on the lower window of
Patient Pressure meter (unit must be operating on battery)
Mode:
Mode:
Mode:
Mode:
Rotary switch to set ventilator operating mode: Standby, Assist/
Control, SIMV, or Pressure Cycle
Low Alarm:
Low Alarm:
Low Alarm:
Low Alarm:
R
otary switch to set limit for Low Pressure alarm: 2-32 cmH
2
O in
increments of 2 cmH
2
O
High Alarm/Limit:
High Alarm/Limit:
High Alarm/Limit:
High Alarm/Limit:
Rotary switch to set limit for High Pressure alarm (Assist/Control,
SIMV modes), or limit for Pressure Cycle mode: 15-90 cmH
2
O in
increments of 5 cmH
2
O
Volume:
Volume:
Volume:
Volume:
Push-to-turn knob to set volume: continuously adjustable from 100
to 220 ml
Breath Rate:
Breath Rate:
Breath Rate:
Breath Rate:
Rotary switch to set breathing rate: 1-20 BPM in increments of one
BPM, and 22-38 BPM in increments of two BPM
Inspiratory Time
Inspiratory Time
Inspiratory Time
Inspiratory Time
R
otary switch to set time for delivery of set volume: 0.5-1.0 sec. in
increments of 0.1 sec.; 1.2 sec.; and 1.5-5.5 sec. in increments of
0.5 sec.
Breathing Effort:
Breathing Effort:
Breathing Effort:
Breathing Effort:
Push-to-turn knob to set pressure level for detecting breathing
effort: continuously adjustable from -10 to +10 cmH
2
O