Check the front panel connection ports for cracks, dents or bent
tubes. Ensure that they are free of foreign matter. LP10 or LP20:
Inspect the
PRESSURE LIMIT
knob to ensure that it is not chipped, bro-
ken, or loose on its shaft. Make sure that the outer ring locks the
small, inner knob.
Rear Panel
Rear Panel
Rear Panel
Rear Panel
Check the rear panel for dirt, or damage. Examine the power cord
for damaged insulation. Check the plug for bent pins or breaks.
Examine the
EXTERNAL BATTERY
,
and
COMMUNICATIONS PORT
for secure
attachment and unbent connector pins.
Inlet Air Filter
Inlet Air Filter
Inlet Air Filter
Inlet Air Filter
Unscrew the cap to the inlet air filter, and check the air filter for
cleanliness. Replace the air filter when it becomes discolored. Insert
a clean filter so any printing is towards the front of the ventilator.
Caution
Caution
Caution
Caution
This filter should be checked weekly (or daily if the ventilator is
being used for transport or in an outdoor environment). A
blocked filter may cause setting error alarms as well as serious
damage to the ventilator.
Caution
Caution
Caution
Caution
Correct any problems found during the visual inspection