Preventive Maintenance
Preventive Maintenance
Preventive Maintenance
Preventive Maintenance
LP6 Plus, LP10, and LP20 Technical Manual
November 2000
Page 5-3
Page 5-3
Page 5-3
Page 5-3
Inlet Filter
Inlet Filter
Inlet Filter
Inlet Filter
Standard Inlet Filter
Standard Inlet Filter
Standard Inlet Filter
Standard Inlet Filter
Frequency
Weekly; daily if used in transit or outdoors
Supplies
❒
replacement filter
Procedure
1.
1.
1.
1.
Twist the threaded plastic cap counterclockwise.
2.
2.
2.
2.
Remove the screens and the old filter.
3.
3.
3.
3.
Wash the screens with soap and water; rinse and dry thoroughly
before reinserting.
4.
4.
4.
4.
Check the filter for dust, dirt or discoloration.
Note:
Filters cannot be cleaned. Replace, if necessary. Any printing on the
replacement filter should face towards the front of the ventilator.
5.
5.
5.
5.
Reassemble in reverse order and secure the plastic filter cap.
Flatpak Inlet Filter
Flatpak Inlet Filter
Flatpak Inlet Filter
Flatpak Inlet Filter
Frequency
Monthly; weekly if used in transit, outdoors, or dusty
conditions
Supplies
❒
replacement Flatpak filter
Procedure
1.
1.
1.
1.
Twist the retainer ring counterclockwise. See figure on the next
page.
2.
2.
2.
2.
Remove and inspect the Flatpak filter for dust, dirt, or discolora-
tion. Filters cannot be cleaned. Replace, if necessary.
3.
3.
3.
3.
Reassemble in reverse order, making sure that the retainer ring
is tight.
Caution
Caution
Caution
Caution
Failure to change the filter may result in serious damage to the
ventilator, and can invalidate the warranty.