30
(2) Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The NSI-III is intended for use in the electromagnetic environment specified below.
The customer or user of the NSI-III should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable
Not applicable
Surge
IEC 61000-4-5
Not applicable
Not applicable
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Not applicable
Not applicable
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3A/m
30A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.