29
ANNEX B
EMC Guidance for Spot Illuminator NSI-III
The Spot Illuminator NSI-III is a medical device. Be sure to install and use it in accordance with the following
EMC (Electromagnetic Compatibility) information.
Be careful that a portable or mobile RF communications equipment may affect the NSI-III. Do not use it
nearby the NSI-III.
⚫
Do not use together with parts other than standard accessories or replacement parts provided by
manufacturer. It may result in increasing emission and decreasing immunity of NSI-III.
⚫
Do not use nearby the electronic equipment or together with it. It may cause malfunction or problem. If it
is necessary to use it, be sure to verify the normal operation under the same condition beforehand.
(1) Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The NSI-III is intended for use in the electromagnetic environment specified below.
The customer or user of the NSI-III should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR11
Group 1
The NSI-III uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR11
Class B
The NSI-III is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable