User Manual / CONVERTISS FOLDING PREMIUM
Made by NAUSICAA Medical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)
18
24
Wooden panels
Damages, wooden chips
Welding
Welding problems
Bed rail bars
Damages, chips
Knobs
Correct fixing
Worn out parts, such as the hinged
points
Damages
II. Electrical measurement in accordance with VDE 0751-1:2001-10 (power from the device, direct )
Proceed as follows:
ok
not
ok
Problem
description
1. Insert the power cord of the bed into the testing outlet of the measuring
device.
2. Connect the probe of the measuring device to a free conductor
element of the frame (such as a screw or similar).
3. Activate the motors with the hand control for the duration of the
measurements
4. Start the measurement procedure with the appropriate device.
Limit
Value
Value of
the first
measure
ment
Current
actual
value
Result: bed SK II (type B)
0,1 ma
ma
ma
Protocol system for testing electrical medical devices
Sheet 2 to 2
III. Functional control
ok
not
ok
Problem
description
Functional control of electrical components
End stroke of the motors
Automatic disconnection
Controller and motors
Unusual noises
Controller and motors
Test according to use instructions
Strain relief cord
Connecting cable properly fixed
Functional control of the mechanical components
Hinged and pivot points
Working smoothly
casters
Brakes, moving forwards
Bed rails
Locking and unlocking
mechanisms
25
Knee raiser
Hinge point
Accessories (ex. Trapeze bar,
handle)
Mounting, damages, suitability
Inspection result:
All values within the allowed range:
yes no
Next inspection date:
The test was passed:
yes no
If the test was not passed:
defective, do not use the bed! => Maintenance
defective, do not use the bed!! => Elimination
The bed does not meet the safety requirements
Mark test applied:
yes no
This inspection protocol included:
Notes:
Tested on:
Examiner:
Signature:
Evaluated on:
Owner/staff:
Signature:
6.3 Spare parts
Replacement parts are available From Apollo Healthcare Technologies Ltd. by indicating
the item number, order number and serial number. The necessary information can be
obtained from the identification label placed on the bed.
Model
Bed model
Type
Article number
Series
Serial number
6. MAINTENANCE