User Manual / CONVERTISS FOLDING PREMIUM

Made by NAUSICAA Medical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)

4. FUNCTIONING

4.1 SPECIAL SAFETY WARNINGS

11

The main control box comes already attached to the 
knee break actuator with all the cables for the 
actuators and the handset connected.

The knee break actuator should be mounted under 
the bed frame so that the motor end (the end where 
the cables connects to the actuator) is towards the 
centre of the bed and the extending ram of the 
actuator is connected to the moving knee break 
section of the mattress platform.

In order to insert the securing pin into the end of the 
extending ram you need to manually lift the knee 
break and pull out the ram so that it is away from the 
control box it is attached to, allowing you to insert the 
locking pin. Please see image to the left.

The backrest actuator should then be mounted under the bed frame again so that the 
motor end is towards the centre of the bed and the extending ram is connected to the 
moving back rest section of the mattress platform.

Ensure all mounting pins are fully inserted through the bed frame mounting points on both 
sides of the actuator and the spring loaded locking loops are in the closed position.

All the cables should be already attached to the control box and numbered 1 – 4 with a 
tag on the opposite end where they connect into each actuator. The number inside the 
control box (re circles above) should correspond with the same number on the other end 
of each cable.

Number 1 cable should be connected into the backrest actuator.
Number 2 cable should be connected into Hi-Lo actuator at the foot end of the bed.
Number 3 cable should be connected into the backrest actuator.
Number 4 cable should be connected into the Hi-Lo actuator at the head end of the bed.   

Pay attention to position the back and foot rest actuators correctly! They should 
never be inverted.

•When making the adjustments make sure that no parts of the body of any patients, users and other 

people, especially children, are positioned under any adjustable elements or under the bed frame to avoid 

being caught and injured.

•Once you have made the necessary adjustments and positioned the bed rails correctly, make sure to lock 

the backrest and leg adjustments possibilities with the hand control, to protect the patient from unwanted 

movements (risk of crushing of his/her limbs).

• In case of use of accessories on the electric bed, make sure such accessories do not crush or cut the 

patient with sharp edges during the legs and backrest adjustments. If this condition is not guaranteed, the 

user must not proceed with the backrest and legs adjustments.

• Make sure the power cable and the hand control cable do not get caught anywhere or get damaged.

4.2 HAND CONTROL

The electric functions of the bed are activated by the patient and by the user with a hand control.

• The electric motors work if the corresponding buttons are pressed.

• It is possible to adjust different sections in both directions.

• The hand control can be attached to the bed where needed with a hook.

• The spiral cable allows freedom of movement.

4.2.1 HAND CONTROL USE

•Before each operation with the bed, make sure the power cable is not stretched, crushed or damaged in 

any way.

•When making the adjustments, make sure there are no obstructions such as furniture or a ceiling inclination, 

to avoid possible damages.

As with all electrical appliances, despite the compliance with prescribed values, it is not possible to totally 

exclude the possibility of interferences from other electrical devices which might be nearby (for ex. the “buzz” 

of a radio). In such a rare case, increase the distance of the bed from the interfering device, do not use the 

same socket or temporarily disconnect such device.

Backrest adjustment / Folding & Unfolding the Bed

When the bed is in use these buttons are used to change the angle of the backrest. These 

buttons are also used to fold and unfold the bed (see Section 3.1 Assembly on page 10 of 

this manual for further details)

Leg support adjustment

When the bed is in use these buttons are used to change the angle of the leg support (knee 

break).

Backrest and leg support combined adjustment (Auto-Contour)

When the bed is in use these buttons are used to change the angle of the backrest and the 

angle of the leg support simultaneously.

12

Make sure not to damage any cables. Verify that the cables are not tangled 
and that they are not able to get caught in any in moving parts during use.

In case of replacement of any individual components, always make sure that 

the electric plugs are then properly inserted into the controller. Only in this way 
will the bed function properly. 

3.1.1 Use of accessories

Apollo Healthcare Technologies can supply various accessories for use with this bed 
including a lifting pole, bed lever and dynamic mattress systems. For correct installation of
accessories please refer to section of this manual.

Use of non-Apollo Healthcare Technologies accessories is not advised. Installing
third party accessories could restrict the moving parts of the bed. The warranty 
does not cover any damages caused by use of third party accessories and your 
warranty could be void if third party accessories are installed and used.

3.2 Operations

Every time before using the bed, the user must make sure that:

the wheels of the bed are locked (see also chapter 4.3).

the power supply is compatible with the bed (230 Volt, 50 Hertz).

the power cable is plugged in and placed in such a way that it cannot be damaged.

the power cable, the motor cables and the hand control cable are not damaged by 
any moving parts of the nursing bed.

there are no obstacles in the way such as bedside furniture or chairs.

all accessories used are installed correctly and do not restrict the free movement of any 
part of the bed especially the backrest and knee break mattress platform.

all control functions  have been accurately  inspected (see also chapter 6.1) and  are 
working properly.

only after carrying out the above operations should the bed be regarded as safe to use.

3.2.1 Requirements of the installation site

Make sure there is enough space to safely allow the whole range of adjustments of the 
bed. There should not be any furniture or other objects close to the bed such as window 
cills.

If the bed is to be used on parquet floor, make sure that the use of wheels will not cause 
any discolouring due to the parquet waterproof treatment. It is possible to use the bed 
on tiles, carpets, laminated floors and linoleum.

A 230 volt power outlet, correctly installed, must be located as near as possible to the
bed.

If there is the need to use other accessories (such as compressor systems, bed levers
etc…), make sure they are correctly fixed and that they work properly. In particular, make 
sure all different cables are safely connected and no moving part of the bed is restricted 
from moving freely as it should. In case of questions or doubts please contact the 
manufacturer of such accessories.

5

2. GENERAL INFORMATION

Before using the bed:

Read through this whole manual in order to avoid causing damages to 

the bed due to improper use.

Before using the nursing bed, the user must verify its perfect functional status and complete 
efficiency. This bed conforms to all the requirements of Directive 93/42/EEC for medical 
products and it is classified as a Class I medical device.

This manual contains safety instructions which must be fully  respected. All the people 
involved with using the bed must know the complete contents of this manual and observe 
the safety warnings listed herein.

Information for the owner:

Every electrical appliance may cause hazards if used incorrectly. Therefore, it is 
necessary to instruct the user on how to use the nursing bed correctly. 

Inform the user of where this manual will be kept.

2.1 Groups of people mentioned in this manual

The manual may refer to the following groups of people that are involved with installing, 
inspecting using, testing or repairing the bed:

Owner
The owner (for example: suppliers of medical products, dealers, health care 
authorities, etc.) can be any physical person or judical body which uses the nursing
bed or  under  whose name the bed is being  used.  The owner is responsible for 
providing the necessary information to the user.

User
the user is the person authorised to use the nursing  bed and  to  carry out any 
operations on the bed on the basis of his/her training, experience or education, and 
who has been trained to use the bed. Furthermore, the user must be able to perform 
basic safety checks on the bed, and also be able to recognise and avoid potential 
hazards, as well as evaluating the condition of the patient.

• Patient

In this manual the term patient refers to the person who needs to be taken care of, 
who might be sick or disabled and therefore in need of using this bed.

Specialised personnel
The term specialised personnel refers to the owners own technicians or personnel 
from the owners appointed company who are authorised, on the basis of their 
training or education, to deliver and assemble the bed.

14

In such a rare case, increase the distance of the bed from the interfering device, do not 
use the same socket or temporarily disconnect such device.

4.2 Hand control

The electric functions of the bed are activated by the patient and by the user with a hand 
control.

The electric motors work if the corresponding buttons are pressed.

It is possible to adjust different sections in both directions.

The hand control can be attached to the bed where needed with a hook.

The spiral cable allows freedom of movement.

Individual functions of the bed can be locked using the hand control.

4.2.1 Hand control use

Backrest adjustment
When the bed is in use these buttons are used to change the angle of 
the backrest.

Leg support adjustment 
When the bed is in use these buttons are used to change the angle of 
the leg support (knee break).

Backrest and leg support combined adjustment (Auto-Contour)
When the bed is in use these buttons are used to change the angle of 
the backrest and the angle of the leg support simultaneously.

Height adjustment of the mattress base 
When the bed is in use these buttons are used to change the height of 
the mattress base. When the height is adjusted there is an automatic 
horizontal positioning of the mattress base in the highest/lowest position.

Angle adjustment of feet – head section (Trendelenburg Tilt)
When the bed is in use these buttons can be used to lower the foot or 
head section independently. When the height is adjusted there is an 
automatic horizontal adjustment of the mattress base if lowered at the 
feet in the highest position and/or in the lowest one.

Before lowering either section, you need to release the brake of the two wheels on the head 
or foot panel end, to avoid possible damages to the floor.

Trendelenburg Tilt - is when the foot end is raised higher than the head end. This is
a potentially dangerous position for some patients. Trendelenburg Tilt should only
be used if appropriate medical advice has been obtained or if you are sure this
position is safe for the patient.

Nurse Lock-Out – Each pair of buttons on the handset can be individually locked out to 
prevent unwanted operation. Using the blue key provided with the handset you can turn 
the central lock to either UNLOCKED (12 o’clock position) or LOCKED (1 o’clock position).

14

In such a rare case, increase the distance of the bed from the interfering device, do not 
use the same socket or temporarily disconnect such device.

4.2 Hand control

The electric functions of the bed are activated by the patient and by the user with a hand 
control.

The electric motors work if the corresponding buttons are pressed.

It is possible to adjust different sections in both directions.

The hand control can be attached to the bed where needed with a hook.

The spiral cable allows freedom of movement.

Individual functions of the bed can be locked using the hand control.

4.2.1 Hand control use

Backrest adjustment
When the bed is in use these buttons are used to change the angle of 
the backrest.

Leg support adjustment 
When the bed is in use these buttons are used to change the angle of 
the leg support (knee break).

Backrest and leg support combined adjustment (Auto-Contour)
When the bed is in use these buttons are used to change the angle of 
the backrest and the angle of the leg support simultaneously.

Height adjustment of the mattress base 
When the bed is in use these buttons are used to change the height of 
the mattress base. When the height is adjusted there is an automatic 
horizontal positioning of the mattress base in the highest/lowest position.

Angle adjustment of feet – head section (Trendelenburg Tilt)
When the bed is in use these buttons can be used to lower the foot or 
head section independently. When the height is adjusted there is an 
automatic horizontal adjustment of the mattress base if lowered at the 
feet in the highest position and/or in the lowest one.

Before lowering either section, you need to release the brake of the two wheels on the head 
or foot panel end, to avoid possible damages to the floor.

Trendelenburg Tilt - is when the foot end is raised higher than the head end. This is
a potentially dangerous position for some patients. Trendelenburg Tilt should only
be used if appropriate medical advice has been obtained or if you are sure this
position is safe for the patient.

Nurse Lock-Out – Each pair of buttons on the handset can be individually locked out to 
prevent unwanted operation. Using the blue key provided with the handset you can turn 
the central lock to either UNLOCKED (12 o’clock position) or LOCKED (1 o’clock position).

14

In such a rare case, increase the distance of the bed from the interfering device, do not 
use the same socket or temporarily disconnect such device.

4.2 Hand control

The electric functions of the bed are activated by the patient and by the user with a hand 
control.

The electric motors work if the corresponding buttons are pressed.

It is possible to adjust different sections in both directions.

The hand control can be attached to the bed where needed with a hook.

The spiral cable allows freedom of movement.

Individual functions of the bed can be locked using the hand control.

4.2.1 Hand control use

Backrest adjustment
When the bed is in use these buttons are used to change the angle of 
the backrest.

Leg support adjustment 
When the bed is in use these buttons are used to change the angle of 
the leg support (knee break).

Backrest and leg support combined adjustment (Auto-Contour)
When the bed is in use these buttons are used to change the angle of 
the backrest and the angle of the leg support simultaneously.

Height adjustment of the mattress base 
When the bed is in use these buttons are used to change the height of 
the mattress base. When the height is adjusted there is an automatic 
horizontal positioning of the mattress base in the highest/lowest position.

Angle adjustment of feet – head section (Trendelenburg Tilt)
When the bed is in use these buttons can be used to lower the foot or 
head section independently. When the height is adjusted there is an 
automatic horizontal adjustment of the mattress base if lowered at the 
feet in the highest position and/or in the lowest one.

Before lowering either section, you need to release the brake of the two wheels on the head 
or foot panel end, to avoid possible damages to the floor.

Trendelenburg Tilt - is when the foot end is raised higher than the head end. This is
a potentially dangerous position for some patients. Trendelenburg Tilt should only
be used if appropriate medical advice has been obtained or if you are sure this
position is safe for the patient.

Nurse Lock-Out – Each pair of buttons on the handset can be individually locked out to 
prevent unwanted operation. Using the blue key provided with the handset you can turn 
the central lock to either UNLOCKED (12 o’clock position) or LOCKED (1 o’clock position).

9

Summary of Contents for CONVERTISS FOLDING PREMIUM

Page 1: ...3485 www nausicaa medical com User Manual CONVERTISS FOLDING PREMIUM Made by NAUSICAA Medical S A S Approved by Ghizlane Labrosse Biomedical Engineer MANUFACTURING STEEL MAXIMUM PATIENT WEIGHT 180 kg CLASS 1 MEDICAL DEVICE ALL OUR DEVICES COMPLYWITH NF EN ISO 60601 2 52 2010 ...

Page 2: ... NAUSICAA Medical S A S Approved by Ghizlane Labrosse Biomedical Engineer 1 HEAD PANEL 2 METAL 4 BAR BED RAILS not shown 3 HAND CONTROL 4 BED RAIL RELEASE LEVER 5 FOOT PANEL 6 WHEELS WITH BRAKE 7 HOUSING FOR LIFTING POLE ON BOTH SIDES 4 5 2 1 3 6 7 ...

Page 3: ...1 Special safety warnings 4 2 Hand control 4 2 1 Hand control use 4 3 Swivel wheels 4 4 Lifting Pole 4 5 Bed Lever 4 6 Dynamic Mattress 4 7 Bed Rails 4 8 Bed End Panels 5 CLEANING AND DISINFECTING 5 1 General information 5 2 Cleaning and disinfection program 5 3 Training of users 5 4 Detergents and disinfectants 5 4 1How to use disinfectants 6 MAINTENANCE 6 1 By the user 6 2 By the owner 6 3 Spare...

Page 4: ...ity of each bed has been extensively tested by the manufacturer in order to avoid any faults This manual provides useful information to the buyer and to the end user on all the functions which will allow a safe and practical daily use of this bed Therefore we recommend you to use this manual as a handy guide and to keep it near your bed In the hope that for you and for the users of this nursing be...

Page 5: ...he nursing bed is a high technology product therefore make sure to read the complete manual when delivered Make sure to use the nursing bed only if it is in good condition The goal of the safety warnings is to avoid injuries to people 2 2 1 SAFETY SYMBOLS USED The following symbols appear in this manual Warnings against injuries This symbol indicates the presence of dangerous power voltage which c...

Page 6: ...of his her training experience or education and who has been trained to use the bed Furthermore the user must be able to perform basic safety checks on the bed and also be able to recognise and avoid potential hazards as well as evaluating the condition of the patient Patient In this manual the term patient refers to the person who needs to be taken care of who might be sick or disabled and theref...

Page 7: ...are made of wood or similar material whose surface is waterproof The surfaces are not harmful in contact with the skin 2 3 4 FRAME The bed is supplied part assembled to be easily transported into the house It is made up of the bed frame with wooden panel inserts head panel and foot panel a mattress base with a motor a pantograph mechanism with an electric actuator two four bar metal bed rails and ...

Page 8: ...ed locking loops are in the closed position All the cables should be already attached to the control box and numbered 1 4 with a tag on the opposite end where they connect into each actuator The number inside the control box re circles above should correspond with the same number on the other end of each cable Number 1 cable should be connected into the backrest actuator Number 2 cable should be c...

Page 9: ...l box re circles above should correspond with the same number on the other end of each cable Number 1 cable should be connected into the backrest actuator Number 2 cable should be connected into Hi Lo actuator at the foot end of the bed Number 3 cable should be connected into the backrest actuator Number 4 cable should be connected into the Hi Lo actuator at the head end of the bed Pay attention t...

Page 10: ...ing third party accessories could restrict the moving parts of the bed The warranty does not cover any damages caused by use of third party accessories and your warranty could be void if third party accessories are installed and used 3 2 Operations Every time before using the bed the user must make sure that the wheels of the bed are locked see also chapter 4 3 the power supply is compatible with ...

Page 11: ...art of the bed is restricted from moving freely as it should In case of questions or doubts please contact the manufacturer of such accessories 2 GENERAL INFORMATION Before using the bed Read through this whole manual in order to avoid causing damages to the bed due to improper use Before using the nursing bed the user must verify its perfect functional status and complete efficiency This bed conf...

Page 12: ...arquet floor make sure that the use of wheels will not cause any discolouring due to the parquet waterproof treatment It is possible to use the bed on tiles carpets laminated floors and linoleum A 230 volt power outlet correctly installed must be located as near as possible to the bed If there is the need to use other accessories such as compressor systems bed levers etc make sure they are correct...

Page 13: ...r make sure that the use of wheels will not cause any discolouring due to the parquet waterproof treatment It is possible to use the bed on tiles carpets laminated floors and linoleum A 230 volt power outlet correctly installed must be located as near as possible to the bed If there is the need to use other accessories such as compressor systems bed levers etc make sure they are correctly fixed an...

Page 14: ...hey should never be inverted Special safety rules for the bed rails The bed rails are there to protect patients from possible accidental falls from the bed However they cannot prevent the patient from voluntarily leaving the bed To make the best use of their intended protection function please observe the following precautions Use only undamaged bed rails and in good conditions positioned at the r...

Page 15: ... liquids which might lead to malfunctioning or damages to the electric components 12 Make sure not to damage any cables Verify that the cables are not tangled and that they are not able to get caught in any in moving parts during use In case of replacement of any individual components always make sure that the electric plugs are then properly inserted into the controller Only in this way will the ...

Page 16: ...y the specialised staff even the normal day to day users such as nurses assistants etc must perform both a visual and a functional inspection before each new use and at regular intervals Recommendation check all mechanical and electrical components once a month In addition check the power cord and the hand control cable after each mechanical stress and after each change of position Therefore it is...

Page 17: ...sed again until the damaged components have been totally replaced or repaired To replace or repair damaged parts please contact the owner of the bed 6 2 By the owner 11 The main control box comes already attached to the knee break actuator with all the cables for the actuators and the handset connected The knee break actuator should be mounted under the bed frame so that the motor end the end wher...

Page 18: ...e cables should be already attached to the control box and numbered 1 4 with a tag on the opposite end where they connect into each actuator The number inside the control box re circles above should correspond with the same number on the other end of each cable Number 1 cable should be connected into the backrest actuator Number 2 cable should be connected into Hi Lo actuator at the foot end of th...

Page 19: ...amages chips Knobs Correct fixing Worn out parts such as the hinged points Damages II Electrical measurement in accordance with VDE 0751 1 2001 10 power from the device direct Proceed as follows ok not ok Problem description 1 Insert the power cord of the bed into the testing outlet of the measuring device 2 Connect the probe of the measuring device to a free conductor element of the frame such as...

Page 20: ...onnecting cable properly fixed Functional control of the mechanical components Hinged and pivot points Working smoothly casters Brakes moving forwards Bed rails Locking and unlocking mechanisms Knee raiser Hinge point Accessories ex Trapeze bar handle Mounting damages suitability Inspection result All values within the allowed range yes no Next inspection date The test was passed yes no If the tes...

Page 21: ...are parts supplied by NAUSICAA Médical Ltd To order spare parts for any requests or assistance and for further information please contact Lits Verticalisateurs Lève Personnes Releveur Mobile Lève Personnes Mobile Téléphone 04 66 71 71 80 Fax 04 66 71 71 81 Mail sav nausicaa medical com 6 4 Electrical components replacement Dangerous for your life life threat due to electric shock Before starting a...

Page 22: ...m the owner and ask for repairs Pressing the keys the movements are activated only for a short time Too much weight on the bed The bed hits against an obstacle Reduce the load Remove the obstacle Movement in one direction only Defective hand control or operating console Inform the owner and ask for repairs Function in contrast with the instruction on the hand control Reversed motor connectors Info...

Page 23: ...vice Main Control Unit Type Linak CA4 Series Control Unit Input voltage AC 230 V 10 50 60 Hz Classification Class II Hand control Type Linak 10 button with individual lock outs Electric motors for height adjustment of the mattress base and backrest and footrest Type Linak LA27 Series Linear Actuators Minimum Average speed at full load mm s 4 5 5 5 8 3 Environmental conditions Noise during adjustme...

Page 24: ...User Manual CONVERTISS FOLDING PREMIUM Made by NAUSICAA Medical S A S Approved by Ghizlane Labrosse Biomedical Engineer 22 POTENCE POUR CONVERTISS 5F et 6F POTC 00001 165 9 LABELLING ...

Page 25: ...rranty does not apply if the claims are consecutive to accident misuse or neglect of the unit by the end customer shipping performed without adequate protection alteration or transformation not validated by NAUSICAA Médical S A S the impact of external elements natural disaster fire shock humidity flood lightning installation and or use in a non compliant way with technical standards and safety in...

Page 26: ...24 User Manual CONVERTISS FOLDING PREMIUM Made by NAUSICAA Medical S A S Approved by Ghizlane Labrosse Biomedical Engineer ...

Page 27: ...25 User Manual CONVERTISS FOLDING PREMIUM Made by NAUSICAA Medical S A S Approved by Ghizlane Labrosse Biomedical Engineer ...

Page 28: ...éléphone 04 66 51 50 80 Fax 04 66 51 50 47 Mail contact nausicaa medical com www nausicaa medical com Votre correspondant Manufactured by NAUSICAA Médical S A S 12 Allée du Piot ZA Pôle Actif 30660 Gallargues le Montueux FRANCE ...

Reviews: