76
Part Number
Type
Guidance
9033UP731/
9033C0731
3 channel amplifier box
-
9031UP742/
9031C0742- *
4 channel amplifier box
-
9031F112-
VEP Monitor
Cable length 5 m
Generic
Color printer
-
Generic
Laser printer
-
9031E022- *
Goggles
Cable length 2.9 m
* Components, items, with high integrity characteristics.
Pins on connectors identified with this
ESD warning symbol should not be touched, and connections should not be
made to these connectors unless ESD precautionary procedures are used.
Precautionary procedures include:
•
Methods to prevent build-up of electrostatic charge (e.g., air conditioning, humidification, conductive floor coverings,
non-synthetic clothing).
•
Discharging one’s body to the frame of the EQUIPMENT or SYSTEM or to earth or a large metal object.
•
Bonding oneself by means of a wrist strap to the EQUIPMENT or SYSTEM or to earth.
Staff that could touch connectors identified with the ESD warning symbol should receive this explanation and training. This
includes clinical/biomedical engineering and health care staff.
ESD training should include an introduction to the physics of electrostatic charge, the voltage levels that can occur in
normal practice and the damage that can be done to electronic components if they are touched by an OPERATOR who is
electrostatically charged. Further, an explanation should be given of methods to prevent build-up of electrostatic charge,
and how and why to discharge one’s body to earth or to the frame of the EQUIPMENT or SYSTEM, or bond oneself by
means of a wrist strap to the EQUIPMENT or SYSTEM or to earth prior to making a connection.
Observe the EN/IEC 60601-1-1 and EN/IEC 60601-1-2 standards in case of connection with other instruments. The connection
of the equipment with other devices is allowed only when the safety requirements for the patient, the user and the
environment are not compromised. If the Manual does not contain adequate information about the possibility of
interconnection with other devices, the user should contact the manufacturer or the nearest authorized servicing center to
obtain information about the effects that coupling devices may have on the patient, the user and the environment.
Radio transmitting equipment, cellular phones, etc., shall not be used in the close proximity of the device since this could
influence the performance of the device.
Portable and mobile RF communications equipment can affect the UltraPro S100. Install and operate the UltraPro S100 according
to the EMC information presented in this section.