Function
Description
5
Spring Retaining Clamp for Power Cord
Place the spring retaining clamp around the power cord to ensure a safe connection.
6
Auxiliary Outlet
-
6 Outlets: 50/60Hz
9031D040x:
100-120V~ total max.
- Continuous Operation 7.4A
- Momentary Operation 12.4A
7.4A equals 850VA, and 12.4A equals 1425VA at 115V~
9031D041x:
200-240V~ total max.
- Continuous Operation: 3.7A
- Momentary Operation: 6.2A
3.7A equals 850VA and 6.2A equals 1425VA at 230V~
7
SN: Device Serial Number
8
REF: Device Reference Number
NOTE
: Unplugging the power line cable from the mains input on the Isolating Transformer disconnects the mains power
to all supplied devices.
Isolation Transformer Classification
Type of protection against electric shock:
•
Class I: Equipment in which protection against electric shock does not rely on basic insulation only, but which
includes an additional safety precaution in that means are provided for the connection of the equipment to the
protective earth conductor in the fixed wiring of the installation in such a way that accessible metal parts cannot
become live in the event of a failure of the basic insulation.
Method(s) of sterilization:
•
Device is not intended to be sterilized.
Degree of patient vicinity:
•
Device suitable for use in patient environment.
Degree of protection against electric shock:
•
Enclosure protectively earthed (treated as applied part Type B).
Degree of protection against harmful ingress of water:
•
IP20: Ordinary equipment (enclosed equipment without protection against ingress of water).
Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide:
•
Equipment
not suitable for use in the presence of such a mixture.
Mode of operation:
•
Continuous operation at normal load and Momentary operation at high load.