81
Notes
The Natus UltraPro S100 system is designed to function under a wide range of environmental conditions without compromise
to Safety or Essential Performance. In the event that an environmental artifact (e.g., ESD, line voltage fluctuations, etc.) is of
sufficient intensity and/ or duration to possibly affect the system performance, the system is designed to enter a fail-safe
mode and temporarily discontinue operation. If this should occur, a message will display notifying the operator that such an
event has occurred. Once the operator has cleared this message, the system will typically reinitialize and indicate that
acquisition can resume with the settings restored to the previous state (other than Stimulation – which must be manually
reset/started).
In severe cases, such as if the Voltage fluctuation dips below 100v for an extended period, the unit may need to be rebooted
to recover. If this type of condition causes persistent issues, please contact your local service representative.