Nicolet V-Series systems
3-4
8/11/16
Warnings
Electric shock
hazard
Connect this equipment only to a three-wire, grounded, hospital-grade receptacle. The three-
connector plug must be inserted into a properly wired three-wire receptacle. If a three-wire
receptacle is not available, a qualified electrician must install one in accordance with the
governing electrical code.
Do not under any circumstances remove the grounding connector from the power plug.
Do not use extension cords or adaptors of any kind. The power cord and plug must be intact
and undamaged.
The appliance coupler (AC Cordset) is the means for disconnecting the V-Series amplifier
from the power mains.
Measure the system’s leakage current after completion of installation, after disconnect/
reconnection of parts of the system, and at regular periodic intervals. For instructions on
how to perform this test please contact Natus Neurology Technical Support. Forward and
reverse polarity: 500 microamperes
The oximeter port is only rated for ESD (Electrostatic Discharge) immunity when
the oximeter is connected. The oximeter should be connected to the oximeter port at
all times when the V Series Amplifier is in use.
Explosion
hazard
Do not use this system in the presence of flammable anesthetics or other flammable
substances.
Summary of Contents for Nicolet V Series
Page 2: ......
Page 5: ...8 11 16 1 1 1 Introduction...
Page 6: ...Nicolet V Series Amplifiers 1 2 August 11 2016 Blank page...
Page 16: ...Nicolet V Series Amplifiers 1 12 August 11 2016 Blank page...
Page 17: ...8 11 16 2 1 2 Product Description...
Page 18: ...Nicolet V Series systems 2 2 8 11 16 Blank page...
Page 26: ...Nicolet V Series systems 2 10 8 11 16 Blank page...
Page 27: ...8 11 16 3 1 3 Installation...
Page 28: ...Nicolet V Series systems 3 2 8 11 16 Blank page...
Page 42: ...Nicolet V Series systems 3 16 8 11 16 Blank page...
Page 43: ...8 11 16 4 1 4 Operation...
Page 44: ...Nicolet V Series Amplifiers 4 2 8 11 16 Blank page...
Page 49: ...8 11 16 5 1 5 Maintenance...
Page 50: ...Nicolet VSeries Amplifiers 5 2 8 11 16 Blank page...
Page 54: ...Nicolet V Series Amplifiers 2 8 11 16 Explosion hazard 3 4 Patient safety 3 5...