Chapter 5
Calibration
5-4
©
National Instruments Corporation
Calibration Programming Considerations
To perform the above-mentioned calibration with the C programming
language, use the NI-DAQ
AO_VWrite and AO_Change_Parameter
APIs. Use the
AO_Change_Parameter
function and the associated
ND_OUTPUT_TYPE
parameter to select either the voltage calibration
channels or the current calibration channels. Use the
AO_VWrite
function
to write the new calibration values to the device. Note that the
AO_Write
function does not support the calibration channels because this
function’s interface cannot specify the required precision. Use
the
Set_DAQ_Device_Info
function and the associated
ND_AO48XDC_SET_POWERUP_STATE
parameter to save the new
calibration channel values to nonvolatile RAM.
Refer to the
NI-DAQ Function Reference Manual for PC Compatibles
for
more information on using these functions to calibrate your
VXI-AO-48XDC module.
To perform the above-mentioned calibration with LabVIEW, use the
AO Update Channel VI and the AO Hardware Config VI. Use the channel
type terminal of the AO Hardware Config VI to select either the voltage
calibration channels or the current calibration channels. Use the AO Update
Channel VI to write the new calibration values to the device. Use the
AO Parameter VI and select the
Save Power-on State
option to save the
new calibration channel values to nonvolatile RAM.