NI VirtualBench Calibration Procedure
|
© National Instruments
|
31
10.
(VB-8012)
Remove the 50 Ω BNC terminator from channel 1 and connect it to channel 2,
as shown in Figure 14.
Figure 14.
VB-8012 Oscilloscope Offset Channel 2 Connection Diagram
11. Repeat steps 2 to 9 for all ranges on channel 2.
Note
When calling
MSO Configure Analog Channel
in step 4, change the
Channel
input to
mso/2
.
12.
(VB-8012)
Remove the 50 Ω BNC terminator from channel 2.
13.
(VB-8034/8054)
Repeat steps 2 to 9 for all ranges on channel 3.
Note
When calling
MSO Configure Analog Channel
in step 4, change the
Channel
input to
mso/3
.
14.
(VB-8034/8054)
Repeat steps 2 to 9 for all ranges on channel 4.
Note
When calling
MSO Configure Analog Channel
in step 4, change the
Channel
input to
mso/4
.
1
VB-8012 Oscilloscope
2
50
Ω
Terminator
1
2