NI VirtualBench Calibration Procedure
Verifying AC Current
To verify the AC current function of the VirtualBench DMM, complete the following steps:
1.
Reset the calibrator.
2.
To verify AC current ranges ≤50 mA, connect the connectors on one end of the low EMF
copper cable to the VirtualBench mA and LO banana plug connectors, and connect the
connectors on the other end of the cable to the HI and LO calibrator binding posts. Figure 6
shows the correct connections.
Figure 6.
Cable Connections for AC Current Ranges
≤
50 mA
3.
Wait 2 minutes for the thermal EMF to stabilize.
4.
Generate 0.0003 A verification point on the calibrator.
5.
Call
DMM Configure Measurement
with the following parameters:
•
Function
=
AC Current
•
Auto Range
=
FALSE
•
Manual Range
=
0.005
6.
Call
DMM Read
. Compare to the limits listed in Table 10.
1
VirtualBench DMM
2
Multifunction Calibrator
3
Low EMF Copper Cable
Table 10.
VirtualBench AC Current Verification Limits
≤
50 mA
Range
(A
rms
)
Verification
Point (A
rms
)
Frequency
(Hz)
1 Year Limits (A
rms
)
24 Hour Limits (A
rms
)
Lower
Upper
Lower
Upper
0.005
0.0003
40
0.0002969
0.0003031
0.0002995
0.0003005
0.0025
40
0.0024925
0.0025075
0.0024989
0.0025011
0.00505
40
0.0050374
0.0050626
0.0050481
0.0050519
0.05
0.003
40
0.002969
0.003031
0.002995
0.003005
0.025
40
0.024925
0.025075
0.024988
0.025012
0.0505
40
0.050374
0.050626
0.050480
0.050520