SCXI-1356 Shielded Cable Installation Guide

Installing the SCXI-1356 Shielded Cable

Perform the following steps to install the SCXI-1356. Refer to Figure 1 an
Figure 2 as you complete the installation.

1.

Power off the VXIbus chassis and the SCXI chassis.

2.

Install the SCXI modules in the SCXI chassis, following the 
instructions in the module user manuals.

3.

Plug the socket connector of the SCXI-1349 into the rear signal 
connector of the SCXI module that will pass analog signals to the 
VXI-MIO Series module.

4.

Secure the SCXI-1349 to the SCXI chassis by threading two of the 
screws through the rear panel of the SCXI-1349 and into the threaded 
strips in the rear of the SCXI chassis.

5.

Connect the MIO-16 branch of the cable to the 68-pin connector of the 
SCXI-1349.

6.

Plug the socket connector of the AI-48/DIO-24 into the rear signal 
connector of the SCXI module or the feedthrough panel to which you 
plan to connect the ACH16–ACH63 signals of the VXI-MIO Series 
module.

7.

Secure the adapter by threading the other two screws through the rear 
panel of the AI-48/DIO-24 and into the threaded strips in the rear of 
the SCXI chassis.

8.

Connect the ACH16–ACH63 branch of the cable to the 68-pin 
connector of the AI-48/DIO-24.

9.

Install the jackscrews and lockwashers as shown in Figure 2.

10. Verify that the two backshell mounting ears are positioned as shown in 

Figure 2. If they are not, remove the mounting ears and install them in 
the correct position.

11. Connect the backshell assembly to the VXI module front connector 

and secure the backshell by tightening both M2.5 mounting screws.

Summary of Contents for SCXI-1356

Page 1: ...00 pin male connector at one end and two 68 pin female connectors on the branched end One 68 pin connector branch is labeled MIO 16 and the other is labeled ACH16 ACH63 Refer to Figure 1 for the parts...

Page 2: ...n introduction to a key concept This font also denotes text that is a placeholder for a word or value that you must supply monospace Text in this font denotes text or characters that you should enter...

Page 3: ...a temporary conductivity caused by condensation must be expected Pollution degree 3 means that conductive pollution occurs or dry nonconductive pollution occurs that becomes conductive due to condensa...

Page 4: ...circuits and circuit breakers Other examples of installation category III are wiring including cables bus bars junction boxes switches socket outlets in the building fixed installation and equipment f...

Page 5: ...CXI chassis 5 Connect the MIO 16 branch of the cable to the 68 pin connector of the SCXI 1349 6 Plug the socket connector of the AI 48 DIO 24 into the rear signal connector of the SCXI module or the f...

Page 6: ...SCXI Modules 2 SCXI Chassis 3 SCXI 1349 Adapter 4 MIO 16 Branch of SCXI 1356 Shielded Cable 5 VXI MIO Series Module 6 96 Pin Male Connector End of SCXI 1356 Shielded Cable 7 ACH16 ACH63 Branch of SCX...

Page 7: ...MIO Series module you use determines which SCXI modules you can use with the SCXI 1356 Connecting the SCXI 1356 to a VXI MIO Series Module The MIO 16 branch of the cable brings the MIO 16 portion of t...

Page 8: ...63 or SCXI 1163HV SCXI 1180 or SCXI 1181 Table 1 lists the pin assignments for connections between the VXI MIO Series module and the MIO 16 branch of the cable Table 1 Cable Connections for the MIO 16...

Page 9: ...O1 B25 17 27 DIO5 C25 51 28 DIO2 A26 49 29 DIO6 B26 16 30 DIO3 C26 47 31 DIO7 A27 48 32 5V A28 8 14 34 35 SCANCLK C27 46 36 DIO4 A25 19 26 DIO1 B25 17 27 DIO5 C25 51 28 DIO2 A26 49 29 DIO6 B26 16 30 D...

Page 10: ...h panel or the SCXI 1181 breadboard module Table 2 lists the pin designations for connections between the VXI MIO Series module and the ACH16 ACH63 cable GPCTR1_OUT B30 40 43 PFI5 UPDATE C30 6 44 PFI6...

Page 11: ...B12 27 15 ACH31 A12 61 16 ACH32 C11 26 17 ACH40 B11 60 18 ACH33 A11 59 19 ACH41 C10 25 20 ACH34 B10 24 21 ACH42 A10 58 22 ACH35 C9 23 23 ACH43 B9 57 24 AISENSE2 A9 56 25 AIGND B17 22 26 ACH36 C8 55 2...

Page 12: ...mode voltage Channel to earth 250 V installation category II Channel to channel 250 V installation category II ACH48 A6 17 35 ACH56 C5 51 36 ACH49 B5 16 37 ACH57 A5 50 38 ACH50 C4 49 39 ACH58 B4 15 40...

Page 13: ...sources include frequently asked questions knowledge bases product specific troubleshooting wizards manuals drivers software updates and more Web support is available through the Technical Support sec...