National Instruments Corporation
3
SCXI-1141/1142/1143 Calibration Procedure
Documentation
The following documents are the primary references for writing your
calibration utility:
•
The
Traditional NI-DAQ Function Reference Help
includes
information about the functions in Traditional NI-DAQ (Legacy).
•
The
DAQ Getting Started Guide
provides instructions for installing
and configuring NI-DAQ devices.
•
The
SCXI Quick Start Guide
includes information for installing and
configuring the SCXI chassis.
•
The
Traditional NI-DAQ User Manual
includes information about
creating applications that use Traditional NI-DAQ (Legacy).
You can add these help files when you install Traditional NI-DAQ
(Legacy). After adding these help files, you can access them by selecting
Start»Programs»National Instruments NI-DAQ»Online-DAQ
Support
. The documents provide instructions about installing and
configuring DAQ devices. The documents also include detailed
information about creating applications that use Traditional NI-DAQ
(Legacy). For more information about the SCXI-1141/1142/1143, refer
to the
SCXI-1141/1142/1143 User Manual
.
Test Equipment
NI recommends using the equipment in Table 1 to verify and calibrate the
SCXI-1141/1142/1143. If these instruments are not available, use the
accuracy requirements listed to select a suitable substitute.
Table 1.
Test Equipment
Equipment
Recommended
Model
Accuracy
Calibrator
Fluke 5700A
50 ppm
DMM
NI 4060
5 1/2 digit, 15 ppm
DAQ Device
NI 6030E
16-bit minimum
Terminal Block
SCXI-1304
N/A