SCXI-1141/1142/1143 Calibration Procedure

10

ni.com

Adjusting the SCXI-1141/1142/1143

This section contains three adjustment procedures: one for measuring gain 
error, one for adjusting calibration gain constants, and one for adjusting 
filter AC gain.

Measuring Gain Errors

Complete the following steps to measure the gain errors in the 
SCXI-1141/1142/1143:

1.

Read the 

Test Conditions

 section in this document.

2.

Refer to Table 9 for the specifications to be tested. Table 9 shows all 
acceptable settings for the module.

3.

Ensure that the E Series DAQ device is connected to the 
SCXI-1141/1142/1143.

4.

Call 

SCXI_Single_Chan_Setup

 to configure the module for 

single-channel measurements. Set the following parameters:

SCXIchassisID

—The value obtained from MAX

moduleSlot

—1

moduleChan

—0

DAQdeviceNumber

—The device number assigned by MAX 

for the E Series DAQ device

5.

Call 

SCXI_Set_Gain

 to configure the module to the gain value you 

want to adjust. Set the following parameters:

SCXIchassisID

—The value obtained from MAX

moduleSlot

—1

moduleChan

—0

gain

—The gain value from Table 9 that you are currently testing

Table 5.  

Passband Test Points for the SCXI-1142

Frequency (kHz)

Lower Limit (V

rms

)

Upper Limit (V

rms

)

2.5

0.9596

0.9889

5.0

0.8909

0.9336

Table 6.  

Passband Test Point for the SCXI-1143

Frequency (kHz)

Lower Limit (V

rms

)

Per Limit (V

rms

)

5.0

0.9882

1.0119

Summary of Contents for SCXI-1141

Page 1: ...e sequence File Page Setup Options directs you to pull down the File menu select the Page Setup item and select Options from the last dialog box This icon denotes a tip which alerts you to advisory in...

Page 2: ...acy which includes high level function calls to simplify the task of writing software to calibrate devices Traditional NI DAQ Legacy supports many programming languages including LabVIEW LabWindows CV...

Page 3: ...install Traditional NI DAQ Legacy After adding these help files you can access them by selecting Start Programs National Instruments NI DAQ Online DAQ Support The documents provide instructions about...

Page 4: ...calling the appropriate calibration functions The steps used in the calibration procedure are as follows 1 Setting up the SCXI 1141 1142 1143 for testing 2 Verifying the existing operation of the SCXI...

Page 5: ...w well the SCXI 1141 1142 1143 is meeting its specifications You can use this information to select the appropriate calibration interval for your application 1 SCXI 1304 Terminal Block 2 SCXI 1141 114...

Page 6: ...d by MAX calOp ND_SELF_CALIBRATE setOfCalConst ND_USER_EEPROM_AREA calRefVolts 0 0 5 Call MIO_Config to enable dithering on the E Series DAQ device measurements Set the following parameters DAQdeviceN...

Page 7: ...arameters SCXIChassisID The device number assigned by MAX moduleSlot 1 channel 0 SCXIgain The gain you are testing TBgain 1 DAQboard The device number for the DAQ device DAQChannel 0 DAQgain 1 numPoin...

Page 8: ...g input channel 0 Refer to Table 7 to determine which pins on the 96 pin front connector correspond to the positive and negative inputs of the specified channel If you are using an SCXI 1304 connected...

Page 9: ...et the calibrator output to a 1 Vrms sine wave at the frequency specified in Tables 4 5 or 6 2 Verify that the DMM output is within the bounds specified by Tables 4 5 or 6 3 Repeat steps 1 and 2 until...

Page 10: ...he SCXI 1141 1142 1143 4 Call SCXI_Single_Chan_Setup to configure the module for single channel measurements Set the following parameters SCXIchassisID The value obtained from MAX moduleSlot 1 moduleC...

Page 11: ...for later use 10 Set the calibrator to the negative test point value for the same gain Skip any input limits that are specified as 0 0 V You need only upper and lower limits for adjustment 11 Read th...

Page 12: ...slot channel The channel you want to adjust opCode 2 calibrationArea 0 rangeCode 0 not used for the SCXI 1141 1142 1143 SCXIgain The gain setting you want to adjust DAQboard The device number assigned...

Page 13: ...the sine wave must be lower than 1 50th of the cutoff frequency For this procedure in sections SCXI 1141 1142 1143 Before Revision F or SCXI 1141 1142 1143 Revision F or Later you will set the filters...

Page 14: ...E Series DAQ device is connected to the SCXI 1141 1142 1143 7 Call SCXI_Set_Gain to configure the module to a gain value of 1 Set the following parameters SCXIchassisID The device number assigned by M...

Page 15: ...channel 0 is pin 4 which is AI 0 11 Set the calibrator voltage to 3 4 Vrms 400 Hz 12 Measure and record the amplitude of the sine wave with the DMM 13 Call SCXI_Configure_Filter to disable the filter...

Page 16: ...nputs of the specified channel If you are using an SCXI 1304 that is connected to the SCXI 1141 1142 1143 connect the calibrator to the AI 0 and AI 0 inputs 6 Connect the DMM to the output of channel...

Page 17: ...MAX moduleSlot 4 channel 0 value The new value decided upon in step 12 14 Repeat steps 11 through 13 until the measured amplitude is as close as possible to the level it was with the filter in bypass...

Page 18: ...3 NC AI 3 23 NC NC NC 22 A GND NC A GND 21 NC NC NC 20 AI 4 NC AI 4 19 NC NC NC 18 AI 5 NC AI 5 17 NC NC NC 16 A GND NC A GND 15 NC NC NC 14 AI 6 NC AI 6 13 NC NC NC 12 AI 7 NC AI 7 11 NC NC NC 10 NC...

Page 19: ...GND AI 3 9 10 A GND AI 4 11 12 A GND AI 5 13 14 A GND AI 6 15 16 A GND AI 7 17 18 A GND NC 19 20 NC NC 21 22 NC NC 23 24 DIG GND SER DAT IN 25 26 SER DAT OUT DAQ D A 27 28 NC SLOT 0 SEL 29 30 NC NC 31...

Page 20: ...41 1142 1143 If the module has been calibrated within the last year the Test Point V value should fall between the Upper Limit V and Lower Limit V values Table 9 SCXI 1141 1142 1143 Specifications Tes...

Reviews: