SCXI-1141/1142/1143 Calibration Procedure
14
ni.com
3.
Remove the side plate of the SCXI chassis.
4.
Install the module into slot 4 of the SCXI chassis.
Figure 3.
Side Plate Removal and Module Installation
5.
6.
Ensure that the E Series DAQ device is connected to the
SCXI-1141/1142/1143.
7.
Call
SCXI_Set_Gain
to configure the module to a gain value of 1.
Set the following parameters:
•
SCXIchassisID
—The device number assigned by MAX
•
moduleSlot
—4
•
moduleChan
—0
•
gain
—1
8.
Call
SCXI_Configure_Filter
to enable the filter bypass. Set the
following parameters:
•
SCXIchassisID
—The value configured in MAX
•
moduleSlot
—40
•
channel
—0
1
Thumbscrews
2
Slot 4
3
Side Plate
®
2
3
M
AIN
FR
AM
E
SC
XI
NA
TIO
NA
L
INS
TR
UME
NT
S
1