RMX-4000 Series Programming Manual
58
1
B
Query Example
:CURR:STAT:REC?
0
A Value is active.
50B
:CURRent:STATic:L1/L2
Channel-Specific
Command
Description
Sets the A/B Value for constant current static
mode, where L1 is A Value and L2 is B Value. The
command is range dependent. If the current range
is Low, the command applies to only the low-
range settings.
Syntax
:CURRent:STATic:L1|L2 <NRf+>[A]
Parameter
<NRf+>[A]
L1 1
Sets A Value to 1 A
L2 2
Sets B Value to 2 A
L1 1A
Sets A Value to 1 A (single
channel only)
L1 MIN
Sets A Value to the
minimum level for the
specific channel
L1 MAX
Sets A Value to the
maximum level for the
specific channel
Example
:CURR:STAT:L1 1
Sets A Value to 1 A for
the current range.
Query Syntax
:CURRent:STATic:L1?/L2? [MAX|MIN]
Return Parameter <NR2> [MAX|MIN]
Current
1 unit = 1 A
Returns the A Value (L1) or
B Value (L2) current
MAX/MIN
Returns the allowable
maximum and minimum