RMX-4000 Series Programming Manual
144
Power Subsystem
:POWer:L1/L2 .......................................................... 144
:POWer:CURRent .................................................... 145
:POWer:RECall ........................................................ 146
:POWer:LOW:AVALue/BVALue .............................. 146
:POWer:LOW:CURRent ........................................... 147
:POWer:HIGH:AVALue/BVALue .............................. 148
:POWer:HIGH:CURRent .......................................... 149
160B
:POWer:L1/L2
Channel-Specific
Command
Description
Sets the A/B Value for constant power mode,
where L1 is A Value and L2 is B Value. The
command is range dependent. If the current range
is Low, the command applies to only the low range
settings.
Syntax
:POWer:L1|L2 <NRf+>[W]
Parameter
<NRf+>[W]
L1 1
Sets A Value to 1 W
L2 2
Sets B Value to 2 W
L1 1W
Sets A Value to 1 W (single
channel only)
L1 MIN
Sets A Value to the
minimum level for the
specific channel
L1 MAX
Sets A Value to the
maximum level for the
specific channel
Example
:POW:L1 1
Sets A Value to 1 W
Query Syntax
:POW:L1?/L2? [MAX|MIN]
Return Parameter <NR2> [MAX|MIN]
Current