Step 4. Setup Synthesizer
1 Insert one end of an SMA cable to the Synthesizer’s RF OUT
SMA connector and the other end of the cable to the applicable
microwave test equipment as shown in Figure 32.
2 Insert the Mini-B end of a USB cable into the Synthesizer’s USB
connector and the A-Type end into the computer’s USB receptacle.
3 Ensure that the power supply is turned off before connecting
DC cables (PN MFSW-ACC03). Verify that the voltage of the
power supply is set b12 V and +12.5 V.
4 Connect the black (ground) wire of the DC bias power cable to
the power supply.
5 Connect the red (+12 V) wire to the power supply.
6 Insert a 20-pin SPI connector into the Synthesizer’s SPI connec-
tor receptacle.
7 Turn on the power supply.
WARNING
Voltage less than +12 V at the Synthesizer connector may result in malfunction,
and voltage in excess of +15 V may damage the Synthesizer.
CAUTION
The QuickSyn Synthesizer generates heat. Therefore, use a heatsink or ensure that
adequate airflow is present to prevent the Synthesizer’s internal temperature from
exc75° C.
NOTE
A quick-start kit that contains all the necessary cables and a power supply to quickly
setup and run the QuickSyn
Synthesizer is available for order as MFSW-ACC06.
NOTE
In addition to a USB interface, other interfaces are available for order as accessories,
including Ethernet, GPIB, and RS232 interfaces.