NI PXIe-6674T Calibration Procedure
b. Set a writable NI-Sync property node to pass FALSE to the
Use PLL?
attribute.
c. Call niSync Connect Clock Terminals VI to connect CLKIN to PXI_CLK10_IN.
d. Call niSync Close VI to close the handle.
LabVIEW Block Diagram
NI-SYNC C Function Call
Call
niSync_SetAttributeViBoolean
with the following parameters:
vi
: “
<
SessionHandle
>
”
terminalName
: “”
attributeID
:
NISYNC_ATTR_CLKIN_USE_PLL
attributeValue
:
VI_FALSE
LabVIEW Block Diagram
NI-SYNC C Function Call
Call
niSync_ConnectClkTerminals
with the following parameters:
vi
: “
<
SessionHandle
>
”
sourceTerminal
:
NISYNC_VAL_CLKIN
destinationTerminal
:
NISYNC_VAL_CLK10
LabVIEW Block Diagram
NI-SYNC C Function Call
Call
niSync_close
with the
following parameter:
vi
: “
<
SessionHandle
>
”