NI PXIe-6674T Calibration Procedure

|

© National Instruments

|

9

b. Call niSync Adjust Oscillator Voltage VI to adjust the voltage of the oscillator.

c. Call niSync Close External Calibration VI to commit the settings and close the 

session.

Reverification

After completing the adjustments to the NI PXIe-6674T, it is important that you verify the 

oscillator frequency operation by repeating the steps listed in the 

Verification

 section. 

Re-verifying after making the adjustments ensures that the NI PXIe-6674T is operating within 

its test limits.

LabVIEW Block Diagram

NI-SYNC C Function Call

Call 

niSync_CalAdjustOscillator

Voltage

 with the following parameters:

vi

: “

<

SessionHandle

>

newVoltage

<

new control voltage

>

oldVoltage

*oldVoltage

LabVIEW Block Diagram

NI-SYNC C Function Call

Call 

niSync_CloseExtCal

 with the 

following parameter:

vi

: “

<

SessionHandle

>

action

:

 

NISYNC_VAL_EXT_CAL_COMMIT

Summary of Contents for PXIe-6674T

Page 1: ...vironments ADEs and programming languages including LabVIEW LabWindows CVI and Microsoft Visual C When you install the drivers you need to install support for only the ADE or programming language you...

Page 2: ...temperature of approximately 23 C 5 C Keep relative humidity below 80 percent Use shielded copper wire for all cable connections to the device Ensure the NI PXIe 6674T has been continuously powered f...

Page 3: ...Installation Guide for details about configuring the PXI Express equipment Verification The following steps outline the procedure for measuring the 10 MHz oscillator frequency on the NI PXIe 6674T and...

Page 4: ...unction Call Call niSync_SetAttributeViBoolean with the following parameters vi SessionHandle terminalName attributeID NISYNC_ATTR_CLKIN_USE_PLL attributeValue VI_FALSE LabVIEW Block Diagram NI SYNC C...

Page 5: ...handle for the device b Call niSync Measure Frequency VI to measure the frequency of the oscillator Table 1 Measurement Duration Device Measure Accuracy Required Measurement Duration NI PXIe 6674T 0...

Page 6: ...limit and high limit range listed under the Calibration Measurement Target Frequency Range in Table 3 the board is considered to be calibrated and no adjustment is needed LabVIEW Block Diagram NI SYNC...

Page 7: ...ives a frequency as close as possible to 10 MHz 1 Complete steps a to c to set the oscillator control voltage The range of acceptable voltage values is 0 0 V to 4 095 V with frequency increasing as vo...

Page 8: ...ency range for your module as shown in Table 3 Adjustment Target Frequency Ranges Adjust the oscillator control voltage up to increase the frequency and down to decrease the frequency 5 Commit the cal...

Page 9: ...e oscillator frequency operation by repeating the steps listed in the Verification section Re verifying after making the adjustments ensures that the NI PXIe 6674T is operating within its test limits...

Page 10: ...ur complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Ap...

Reviews: