NI PXIe-6674T Calibration Procedure

Calibration Interval

The measurement requirements of the application determine how often you should calibrate the 

device to ensure its accuracy. National Instruments recommends that you calibrate the 

NI PXIe-6674T 

yearly

. You may want to shorten this interval based on the application demands.

Password

The default password for password-protected operations is 

NI

.

Test Equipment

Calibrating the NI PXIe-6674T requires the following equipment.

Test Conditions

Install the NI PXIe-6674T in the system timing slot of the PXI Express chassis. The 

calibration procedure requires features of PXI Express that are accessible only in the system 

timing slot.

Maintain a temperature of approximately 23 °C ± 5 °C.

Keep relative humidity below 80 percent.

Use shielded copper wire for all cable connections to the device.

Ensure the NI PXIe-6674T has been continuously powered for at least 24 hours prior to 

calibration to allow the 10 MHz oscillator frequency to stabilize.

Keep PXI Express chassis filters clean and fan speed set to High.

Ensure the frequency reference has been continuously powered, as specified by its manual, 

for its accuracy to be within 0.75 parts-per-billion (ppb) prior to calibration.

Calibration Procedure

The steps used in the calibration procedure are as follows:
1. Initial setup.
2. Verification.

Equipment

Recommended Model

Requirements

10 MHz clock source

Symmetricom/Datum 8040

Accurate to within 

0.75 parts-per-billion (ppb)

BNC-SMA cable

PXI Express chassis

NI PXIe-1062Q, NI PXIe-1075, 

NI PXIe-1082, NI PXIe-1085, 

NI PXIe-1086

Must have a system timing slot

Summary of Contents for PXIe-6674T

Page 1: ...vironments ADEs and programming languages including LabVIEW LabWindows CVI and Microsoft Visual C When you install the drivers you need to install support for only the ADE or programming language you...

Page 2: ...temperature of approximately 23 C 5 C Keep relative humidity below 80 percent Use shielded copper wire for all cable connections to the device Ensure the NI PXIe 6674T has been continuously powered f...

Page 3: ...Installation Guide for details about configuring the PXI Express equipment Verification The following steps outline the procedure for measuring the 10 MHz oscillator frequency on the NI PXIe 6674T and...

Page 4: ...unction Call Call niSync_SetAttributeViBoolean with the following parameters vi SessionHandle terminalName attributeID NISYNC_ATTR_CLKIN_USE_PLL attributeValue VI_FALSE LabVIEW Block Diagram NI SYNC C...

Page 5: ...handle for the device b Call niSync Measure Frequency VI to measure the frequency of the oscillator Table 1 Measurement Duration Device Measure Accuracy Required Measurement Duration NI PXIe 6674T 0...

Page 6: ...limit and high limit range listed under the Calibration Measurement Target Frequency Range in Table 3 the board is considered to be calibrated and no adjustment is needed LabVIEW Block Diagram NI SYNC...

Page 7: ...ives a frequency as close as possible to 10 MHz 1 Complete steps a to c to set the oscillator control voltage The range of acceptable voltage values is 0 0 V to 4 095 V with frequency increasing as vo...

Page 8: ...ency range for your module as shown in Table 3 Adjustment Target Frequency Ranges Adjust the oscillator control voltage up to increase the frequency and down to decrease the frequency 5 Commit the cal...

Page 9: ...e oscillator frequency operation by repeating the steps listed in the Verification section Re verifying after making the adjustments ensures that the NI PXIe 6674T is operating within its test limits...

Page 10: ...ur complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Ap...

Reviews: