NI PXIe-6674T Calibration Procedure

|

© National Instruments

|

3

3. Adjustment.
4. Reverification.

Initial Setup

Complete the following steps to set up the NI PXIe-6674T for calibration.
1. Make sure the PXI Express chassis involved in the calibration procedure is powered off.
2. Install the NI PXIe-6674T board into the system timing slot of your chassis.

Note

The module 

must

 be installed in the system timing slot of the chassis.

3. Power on the PXI Express chassis.
4. Make sure that all the appropriate driver and application software is installed on the host 

computer.

5. Configure the hardware properly with Measurement & Automation Explorer (MAX). Refer 

to the 

NI PXIe-6674T Installation Guide

 for details about configuring the PXI Express 

equipment.

Verification

The following steps outline the procedure for measuring the 10 MHz oscillator frequency on the 

NI PXIe-6674T and determining whether the device requires adjustment to meet the published 

specifications.

Connecting the Clock Source

1. Connect the external 10 MHz reference source to the CLKIN connector on the 

NI PXIe-6674T.

2. Program the NI PXIe-6674T to route the CLKIN signal to PXI_CLK10_IN without using 

its 10 MHz PLL by completing steps a through d.

Note

Use the data in the C function call reference as inputs to your LabVIEW VI 

where applicable.

a. Call niSync Initialize VI to set up a handle for the device.

LabVIEW Block Diagram

NI-SYNC C Function Call

Call 

niSync_init

 with the following 

parameters:

resourceName

Dev1

idQuery

NULL

resetDevice

VI_TRUE

vi

*SessionHandle

Summary of Contents for PXIe-6674T

Page 1: ...vironments ADEs and programming languages including LabVIEW LabWindows CVI and Microsoft Visual C When you install the drivers you need to install support for only the ADE or programming language you...

Page 2: ...temperature of approximately 23 C 5 C Keep relative humidity below 80 percent Use shielded copper wire for all cable connections to the device Ensure the NI PXIe 6674T has been continuously powered f...

Page 3: ...Installation Guide for details about configuring the PXI Express equipment Verification The following steps outline the procedure for measuring the 10 MHz oscillator frequency on the NI PXIe 6674T and...

Page 4: ...unction Call Call niSync_SetAttributeViBoolean with the following parameters vi SessionHandle terminalName attributeID NISYNC_ATTR_CLKIN_USE_PLL attributeValue VI_FALSE LabVIEW Block Diagram NI SYNC C...

Page 5: ...handle for the device b Call niSync Measure Frequency VI to measure the frequency of the oscillator Table 1 Measurement Duration Device Measure Accuracy Required Measurement Duration NI PXIe 6674T 0...

Page 6: ...limit and high limit range listed under the Calibration Measurement Target Frequency Range in Table 3 the board is considered to be calibrated and no adjustment is needed LabVIEW Block Diagram NI SYNC...

Page 7: ...ives a frequency as close as possible to 10 MHz 1 Complete steps a to c to set the oscillator control voltage The range of acceptable voltage values is 0 0 V to 4 095 V with frequency increasing as vo...

Page 8: ...ency range for your module as shown in Table 3 Adjustment Target Frequency Ranges Adjust the oscillator control voltage up to increase the frequency and down to decrease the frequency 5 Commit the cal...

Page 9: ...e oscillator frequency operation by repeating the steps listed in the Verification section Re verifying after making the adjustments ensures that the NI PXIe 6674T is operating within its test limits...

Page 10: ...ur complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Ap...

Reviews: