Figure 6. Sources and Destinations for PXIe-6674 Signal Routing Operations
Using Front Panel PFIs as Single Ended Inputs
The front-panel PFIs can receive external signals from 0 to +5 V. They can be terminated
programmatically with 50 Ω resistances to match the cable impedance and minimize
reflections.
Note
Terminating the signals with a 50 Ω resistance is recommended when the
source is another PXIe-6674 or any other source with a 50 Ω output.
The voltage thresholds for the front-panel PFI inputs are programmable. The input signal is
generated by comparing the input voltage on the PFI connectors to the voltage output of
software-programmable DACs. The thresholds for the PFI lines are individually
PXIe-6674 User Manual
|
© National Instruments
|
17