PXIe-5665 Calibration Procedure
|
© National Instruments
|
15
Characterizing RF Source Power (with Lowpass Filter)
Note
Prior to starting this procedure, zero and calibrate the power sensor according
Zeroing and Calibrating the Power Sensor
section of this document.
1.
Connect the RF source 2 output to power sensor B through the anti-distortion test fixture.
The completed equipment setup is shown in Figure 3.
Figure 3.
RF Source Power (with Lowpass Filter) Characterization Equipment Setup
2.
Disable the RF source 1 output.
1
RF Source 1
2
RF Source 2
3
SMA (m)-to-SMA (m) Cable
4
SMA (m)-to-SMA (m) Cable
5
Anti-Distortion Test Fixture
6
SMA (m)-to-SMA (m) Cable
7
3.5 mm (f)-to-3.5 mm (f) Adaptor
8
Power Sensor B
9
Power Meter
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET 166375A
RF
INPUT 2
6
1
5
2
8
7
4
3
9