PXIe-5665 Calibration Procedure
|
© National Instruments
|
41
Figure 14.
PXIe-5665 14 GHz VSA Non-Input-Related Spurs Verification
Equipment Setup
2.
Create a new session for the PXIe-5665.
3.
Configure the PXIe-5665 according to the following property settings:
•
Acquisition Type: Spectrum
•
Averaging Mode: RMS Averaging
•
Number of Averages: 4
•
Ref Clock Source: PXI_Clk
•
Ref Clock Rate: 10 MHz
•
Channel Coupling: AC Coupled
•
Preamp Enabled: Disabled
•
Span: 50 MHz (25 MHz for the PXIe-5665 with 25 MHz bandwidth)
•
Device Instantaneous Bandwidth: 50 MHz (25 MHz for the PXIe-5665 with 25 MHz
bandwidth)
•
Resolution Bandwidth: 2 kHz
•
RF Attenuation: 0 dB
•
Reference Level: -60 dBm
•
FFT Window Type: Flat Top
•
(PXIe-5665 14 GHz VSA)
Preselector Enabled: Disabled
1
50
Ω
Termination
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD
SENSITIVE
OUT
ALL PORTS
50
Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD
SENSITIVE
TTL
6.3 Vp-p
MAX
2 Vp-p
NOM
+20 dBm MAX
50
Ω
50
Ω
50
Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD
SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50
Ω
+15 dBm MAX
25 V DC MAX
+15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC
+20 dBm MAX > 10 MHz
+10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J