PXIe-5665 Calibration Procedure
|
© National Instruments
|
59
5.
Disable the RF source 1 and RF source 2 outputs.
Note
When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6.
Set the RF source 1 power to the
RF Source 1 Programmed Power
1
corresponding to one
of the following values:
•
Preamplifier disabled: -10 dBm
•
Preamplifier enabled: -30 dBm
7.
Set the RF source 2 power to the
RF Source 2 Programmed Power
2
corresponding to one
of the following values:
•
Preamplifier disabled: -10 dBm
•
Preamplifier enabled: -30 dBm
8.
Configure the PXIe-5665 Reference Level to one the following values:
•
Preamplifier disabled: -10 dBm
•
Preamplifier enabled: -30 dBm
9.
For each test frequency from Table 32 or Table 33, calculate the following four frequencies:
•
IMD Low Frequency
= Test Frequency - 1.05 MHz
•
P1 Frequency
= Test Fre 350 kHz
•
P2 Frequency
= Test Frequency - 350 kHz
•
IMD High Frequency
= Test Fre 1.05 MHz
1
The RF Source 1 Programmed Power
value was measured in the
Characterizing RF Source Power
(Combined) for Frequencies >700 MHz
equipment characterization procedure.
2
The
RF Source 2 Programmed Power
value was measured in the
Characterizing RF Source Power
(Combined) for Frequencies >700 MHz
equipment characterization procedure.
Table 32.
PXIe-5665 3.6 GHz VSA TOI (Frequencies >700 MHz)
Verification Test Frequencies
Start Frequency
Stop Frequency
Step Size
800 MHz
3.3 GHz
500 MHz
3.59895 GHz
—
—