![background image](http://html1.mh-extra.com/html/national-instruments/pxie-4147/pxie-4147_getting-started-manual_3605853013.webp)
Operating System
Description
Windows 10/8.1
Right-click the Start button, and select
Device Manager
.
Windows 7
Select
Start
»
Control Panel
»
Device Manager
.
5.
Verify the PXIe-4147 appears in the Device Manager.
a)
Under an NI entry, confirm that a PXIe-4147 entry appears.
Note
If you are using a PC with a device for PXI remote control system,
under
System Devices
, also confirm that no error conditions appear for the
PCI-to-PCI Bridge
.
b)
If error conditions appear, reinstall NI-DCPower and the PXIe-4147.
Why Is the ACCESS LED Off When the Chassis Is
On?
Before proceeding, verify that the PXIe-4147 appears in MAX.
If the ACCESS LED fails to light after you power on the chassis, a problem may exist with the
chassis power rails, a hardware module, or the LED.
Notice
Apply external signals only while the PXIe-4147 is powered on. Applying
external signals while the module is powered off may cause damage.
1.
Disconnect any signals from the module front panels.
2.
Power off the chassis.
3.
Remove the module from the chassis and inspect it for damage. Do not reinstall a
damaged module.
4.
Install the module in a different chassis slot from which you removed it.
5.
Power on the chassis.
Note
If you are using a PC with a device for PXI remote control system,
power on the chassis before powering on the computer.
6.
Verify that the module appears in MAX.
7.
Reset the module in MAX and perform a self-test.
Related Information
What Should I Do if the PXIe-4147 Fails the Self-Test?
1.
Restart the system.