2.
Launch MAX.
•
Failed self-test
1.
Perform self-calibration, then perform the self-test again. The PXIe-4147 must
be calibrated to pass the self-test.
•
Failed self-calibration
1.
Perform self-calibration again.
3.
Power off the chassis.
4.
Reinstall the failed module in a different slot.
5.
Power on the chassis.
6.
Perform the self-test again.
Where To Go Next
custom applications with
an application programming
interface (API).
InstrumentStudio
NI-DCPower Instrument Driver
about hardware features
or review device
specifications.
more about your products through ni.com.
the application
development environment (ADE)
for your application.
Getting Started
with LabVIEW
Getting Started with
LabWindows/CVI
NI DC Power Supplies
and SMUs Help*
NI DC Power Supplies
and SMUs Help*
NI-DCPower Examples*
*This item is also installed with the driver software.
Located online at ni.com/manuals
Located using the NI Example Finder
EXPLORE
LEARN
CREATE
DISCOVER
Support
ni.com/support
Services
ni.com/services
PXIe-4147 Specifications
NI Community
ni.com/community
Power Supplies
Solutions
ni.com/powersupplies
NI PXIe-4147
4-CH PRECISION SMU
Worldwide Support and Services
The NI website is your complete resource for technical support. At
, you have
access to everything from troubleshooting and application development self-help resources to
email and phone assistance from NI Application Engineers.
Visit
for information about the services NI offers.
14
|
ni.com
|
PXIe-4147 Getting Started Guide