Optional Items

PXI Slot Blocker Kit (NI part number 199198-01)

Notice

 

The PXIe-4147 provides higher maximum current and more precise

current measurement capabilities than earlier PXIe-414

x

 modules.

Consequently, PXIe-4147 performance can be compromised by the accessories
associated with those modules. The following accessories are not recommended
for use with the PXIe-4147.

DB25F-DB25F, 25-pin D-SUB Low Leakage Cable for SMU

1 m (NI part number 782015-01)

2 m (NI part number 782015-02)

Screw Terminal Connector Kit for PXIe-414x SMUs (NI part number
781974-01)

Visit 

ni.com

 for more information about these additional items.

Preparing the Environment

Ensure that the environment you are using the PXIe-4147 in meets the following
specifications:

Temperature and Humidity

Temperature

Operating

0 °C to 55 °C

1

Storage

-40 °C to 71 °C

Humidity

Operating

10% to 90%, noncondensing

2

Storage

5% to 95%, noncondensing

1

Not all chassis can achieve this ambient temperature range. Refer to PXI chassis specifications to
determine the ambient temperature ranges your chassis can achieve.

2

When transitioning a device from a storage or operation environment with relative humidity above
70%, device should be allowed to stabilize in the lower humidity environment for several hours
before use. Refer to the PXIe-4147 

Programming and Measurement Accuracy/Resolution

specifications for additional performance derating when operating above 70% relative humidity.

PXIe-4147 Getting Started Guide

  | 

© National Instruments

  | 

3

Summary of Contents for PXIe-4147

Page 1: ...f Calibrating the PXIe 4147 9 Programming the PXIe 4147 10 Troubleshooting 12 What Should I Do if the PXIe 4147 Doesn t Appear in MAX 12 Why Is the ACCESS LED Off When the Chassis Is On 13 What Should...

Page 2: ...omponents or other signs of damage Notice Never touch the exposed pins of connectors Note Do not install a module if it appears damaged in any way 3 Unpack any other items and documentation from the k...

Page 3: ...nt Ensure that the environment you are using the PXIe 4147 in meets the following specifications Temperature and Humidity Temperature Operating 0 C to 55 C1 Storage 40 C to 71 C Humidity Operating 10...

Page 4: ...ver support for the PXIe 4147 was first available in NI DCPower 20 0 5 Follow the instructions in the installation prompts Note Windows users may see access and security messages during installation A...

Page 5: ...Express Peripheral Slot PXIe 4147 modules can be placed in PXI Express peripheral slots PXI Express hybrid peripheral slots or PXI Express system timing slots 7 Touch any metal part of the chassis to...

Page 6: ...over all empty slots using EMC filler panels or fill using slot blockers to maximize cooling air flow depending on your application 13 Power on the chassis Related Information Why Is the ACCESS LED Of...

Page 7: ...H 1 Sense LO Pin 7 CH 2 Guard Pin 8 CH 2 Sense HI Pin 9 CH 2 Sense LO Pin 10 CH 3 Guard Pin 11 CH 3 Sense HI Pin 12 CH 3 Sense LO Pin 13 N C Pin 14 CH 0 Output HI Pin 15 CH 0 Guard Pin 16 CH 0 Output...

Page 8: ...or is in error due to the voltage generated or measured by the device Refer to the driver software for possible sources The device will not operate until the error is cleared and or the device is res...

Page 9: ...e self calibration after the first time you install the PXIe 4147 1 Install the PXIe 4147 and let it warm up for 30 minutes Note Warm up begins when the PXI Express chassis has been powered on and the...

Page 10: ...Ie 4147 You can generate signals interactively using InstrumentStudio or you can use the NI DCPower instrument driver to program your device in the supported ADE of your choice 10 ni com PXIe 4147 Get...

Page 11: ...io year This launches InstrumentStudio and runs a soft front panel populated with devices detected on your system From Measurement Automation Explorer MAX select a device and then click Test Panels Th...

Page 12: ...nstruments folder C C Available at Program Files IVI Foundation IVI Refer to the Creating an Application with NI DCPower in Microsoft Visual C and C topic of the NI DC Power Supplies and SMUs Help to...

Page 13: ...xist with the chassis power rails a hardware module or the LED Notice Apply external signals only while the PXIe 4147 is powered on Applying external signals while the module is powered off may cause...

Page 14: ...VIEW Getting Started with LabWindows CVI NI DC Power Supplies and SMUs Help NI DC Power Supplies and SMUs Help NI DCPower Examples This item is also installed with the driver software Located online a...

Page 15: ...ort and ensures that you receive important information updates from NI NI corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 USA For up to date contact informati...

Page 16: ...ents EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy a...

Reviews: