Chapter 2
Connecting Signals
2-4
ni.com
Quarter-Bridge Type I
This section provides information for connecting the quarter-bridge
strain-gauge configuration type I. Figure 2-3 shows the quarter-bridge
type I circuit wiring diagram. Refer to the
of Chapter 4,
, for more information.
Note
If you have a two-wire strain-gauge element, connect the active element leads to
S+ (pin 1) and P+ (pin 2) and short S+ (pin 1) to QTR/SCB (pin 9) and SCA (pin 4).
Figure 2-3.
Quarter-Bridge I Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•
R
1
and R
2
are half-bridge completion resistors.
•
R
3
is the quarter-bridge completion resistor (dummy resistor).
•
R
4
is the active element measuring tensile strain (+
ε
).
•
V
EX
is the excitation voltage.
•
R
L
is the lead resistance.
•
V
CH
is the measured voltage.
•
R
S
is shunt calibration resistor A.
Note
The value of the quarter-bridge completion resistor (dummy resistor) must equal the
nominal resistance of the strain gauge. NI recommends using a 0.1% precision resistor.
If a less precise resistor is used, offset null compensation calibration is not as accurate.
Refer to the
section of Chapter 4,
, for more
information.
+
–
+
–
R
L
NI PXI-4220
Transducer
R
1
R
2
R
4
R
3
V
CH
S+ (Pin 1)
P+ (Pin 2)
P– (Pin 7)
Shunt
Cal A
R
L
R
S
SCA (Pin 4)
SCCOM (Pin 5)
QTR/SCB (Pin 9)
V
EX