Full-Bridge Type II
This section provides information for connecting the full-bridge
strain-gauge configuration type II. Figure 2-8 shows the full-bridge type II
circuit wiring diagram. Refer to the
section of
Chapter 4,
, for more information.
Figure 2-8.
Full-Bridge Type II Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•
R
1
is an active element measuring compressive Poisson effect (–
νε
).
•
R
2
is an active element measuring tensile Poisson effect (+
νε
).
•
R
3
is an active element measuring compressive strain (–
ε
).
•
R
4
is an active element measuring tensile strain (+
ε
).
•
V
EX
is the excitation voltage.
•
R
L
is the lead resistance.
•
V
CH
is the measured voltage.
•
R
S
is shunt calibration resistor A.
+
–
+
–
Shunt
Cal A
R
L
R
L
NI PXI-4220
Transducer
R
1
R
2
R
4
R
3
V
CH
R
L
R
L
S– (Pin 6)
R
S
V
EX
S+ (Pin 1)
P+ (Pin 2)
P– (Pin 7)
SCA (Pin 4)
SCCOM (Pin 5)