![National Instruments NI PXI-4220 User Manual Download Page 97](http://html.mh-extra.com/html/national-instruments/ni-pxi-4220/ni-pxi-4220_user-manual_709323097.webp)
Appendix A
Specifications
A-6
ni.com
Regulation, no load to 120
Ω
load
With remote sense ...........................±0.003%
Without remote sense ......................±0.30%
Temperature drift ....................................±0.005%/°C, ±30
µ
V/°C max
Noise, DC to 10 kHz...............................200
µ
V
pp
Remote sense ..........................................Error less than ±0.02% per
Ω
of
lead resistance, both leads
Protection................................................Surge arrestors in parallel with
excitation terminals, shunt to
ground
Bridge Completion
1
Half-bridge..............................................Two precision resistors,
5 k
Ω
each
Quarter-bridge ........................................350
Ω
resistor socketed on
PXI-4220 (120
Ω
resistor
included in kit)
Shunt Calibration
2
Type ........................................................Two independent points
Resistor ...................................................100 k
Ω
resistors socketed on
PXI-4220
Switch resistance ....................................32
Ω
typical, 50
Ω
max
Switch off leakage ..................................<1 nA
Switch break-down voltage ....................±60 V
1
Half-bridge completion is performed inside the module and configured under software control. The quarter-bridge completion
resistor is located near the front of the PXI-4220 and is socketed. Resistors shipped with the PXI-4220 are 350
Ω
(installed)
and 120
Ω
(in bag with kit) RN-55 style (standard 1/4 W size). The tolerance is ±0.1% and the temperature coefficient is
10 ppm/°C max.
2
Shunt calibration resistors are located near the front of the PXI-4220 and are socketed. Resistors shipped with the PXI-4220
are 100 k
Ω