Assuming typical thermal EMF, the total error in
thermocouple measurement at 200 °C for the
NI PXI-2527/TB-2627 with a K-type thermocouple
is ±2.76 °C.
Physical Characteristics
Relay type........................................Electromechanical,
latching
Relay contact material .....................Palladium-ruthenium,
gold covered
I/O connector ...................................100 position HDI right
angle, male
PXI power requirement ...................6 W at 5 V,
2.5 W at 3.3 V
Dimensions (L
×
W
×
H).................3U, one slot,
PXI/cPCI module
21.6
×
2.0
×
13.0 cm
(8.5
×
0.8
×
5.1 in.)
Weight .............................................209 g (7.4 oz)
Environment
Operating temperature .....................0 °C to 55 °C
Note
If you are using the NI PXI-2527 with the
NI PXI-101
X
or the NI PXI-1000B chassis,
operating temperature for the NI PXI-2527 is
0 °C to 45 °C.
Note
Do
not
operate the NI PXI-2527 above the
maximum operating temperature of the chassis.
Storage temperature.........................–20 °C to 70 °C
Relative humidity ............................5% to 85%,
noncondensing
Pollution Degree..............................2
Maximum altitude ...........................2,000 m
Indoor use only.
Shock and Vibration
Operational Shock ...........................30 g peak, half-sine,
11 ms pulse (Tested
in accordance with
IEC 60068-2-27. Test
profile developed in
accordance with
MIL-PRF-28800F.)
Random Vibration
Operating .................................... 5 to 500 Hz, 0.3 g
rms
Nonoperating .............................. 5 to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC 60068-2-64.
Nonoperating test profile
exceeds the requirements
of MIL-PRF-28800F,
Class 3.)
Accessories
Visit
ni.com
for more information about NI accessories.
Caution
You
must
install mating connectors
according to local safety codes and standards and
according to the specifications provided by the
connector manufacturer. You are responsible for
verifying safety compliance of third-party
connectors and their usage according to the relevant
standard(s), including UL and CSA in North
America and IEC and VDE in Europe.
Figure 2 shows the NI PXI-2527 power-on state.
Table 1.
Accessories for the NI PXI-2527
Accessory
Manufacturer
Part Number
NI TB-2627 terminal
block
NI
779358-01
IM42GR relay
replacement kit
NI
779356-01
100-pin HDI right
angle mating
connector
*
NI
779311-01
Mating front panel
connector, vertical
*
AMP
533285-1
Mating front panel
connector,
right-angle
*
AMP
532903-2
*
PCB mount, additional cover or enclosure required.