background image

10

|

ni.com

|

NI 9229/9239 Calibration Procedure

g.

Set the calibrator to Standby mode (STBY).

h.

Repeat steps d through g for each calibration voltage in the array.

i.

Close the calibration session.

4.

Disconnect the NI 9229/9239 from the calibrator.

EEPROM Update

When an adjustment procedure is completed, the NI 9229/9239 internal calibration memory 
(EEPROM) is immediately updated.

If you do not want to perform an adjustment, you can update the calibration date and onboard 
calibration temperature without making any adjustments by initializing an external calibration

setting the C Series calibration temperature, and closing the external calibration.

Reverification

Repeat the 

Verification

 section to determine the As-Left status of the device.

Note

If any test fails Reverification after performing an adjustment, verify that you 

have met the 

Test Conditions

 before returning your device to NI. Refer to 

Worldwide 

Support and Services

 for assistance in returning the device to NI.

Accuracy Under Calibration Conditions

The following calibration specifications are valid under the following conditions:

Ambient temperature 23 ± 5 °C

NI 9229/9239 installed in slot 8 of an NI cDAQ-9178 chassis

Slots 1 through 7 of the NI cDAQ-9178 chassis are empty

Note

The test limits listed in Table 4 are derived using the values in Table 14.

Table 13.  

Adjustment Configuration

Physical Channel

Reference Value

cDAQMod8/ai

x

The reference value from the array of adjustment points

Table 14.  

NI 9229/9239 Accuracy Under Calibration Conditions

Module

Gain Error

Offset Error

NI 9229

±0.04%

±7.66 mV

NI 9239

±0.04%

±1.35 mV

Summary of Contents for NI 9229

Page 1: ...y Verification 4 Gain Matching Verification 6 Phase Matching Verification 7 Adjustment 9 Accuracy Adjustment 9 EEPROM Update 10 Reverification 10 Accuracy Under Calibration Conditions 10 Worldwide Sup...

Page 2: ...nstructions and Specifications NI 9229 9239 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabV...

Page 3: ...ctions to the device Use twisted pairs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C Keep relative humidity below 80 Allow a warm up time of at least 10 minut...

Page 4: ...elf Test to ensure that the device is working properly Verification The following performance verification procedures describe the sequence of operation and provide test points required to verify the...

Page 5: ...Accuracy Verification Module Physical Channels Min V Max V Scaled Units Terminal Configuration 9229 AI0 3 60 60 Volts Differential 9239 AI0 3 10 10 Volts Differential Table 3 NI 9229 9239 Timing Conf...

Page 6: ...ure function generator according to Table 6 3 Acquire and average samples a Create and configure an AI voltage task on the NI 9229 9239 according to Table 7 Table 6 Function Generator Configuration fo...

Page 7: ...If the Gain Matching verification procedure determines that the NI 9229 9239 is outside of the limits refer to Worldwide Support and Services for assistance in returning the device to NI Phase Matchin...

Page 8: ...able the function generator g Clear the task 5 Calculate the phase for each channel on the NI 9229 9239 NI recommends using the Extract Single Tone Information VI to calculate phase Table 9 Function G...

Page 9: ...r as shown in Figure 1 Note If the calibrator outputs are truly floating the negative output must be connected to a quiet earth ground as well as AI to give the entire system a ground reference 3 Adju...

Page 10: ...determine the As Left status of the device Note If any test fails Reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to...

Page 11: ...om manuals Worldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and appl...

Reviews: