NI 9229/9239 Calibration Procedure

|

© National Instruments

|

3

Test Equipment

Table lists the equipment recommended for the performance verification and adjustment 
procedures. If the recommended equipment is not available, select a substitute using the 
requirements listed in Table 1.

Test Conditions

The following setup and environmental conditions are required to ensure the NI 9229/9239

 

meets published specifications.

Keep connections to the device as short as possible. Long cables and wires act as antennas, 
picking up extra noise that can affect measurements.

Verify that all connections to the NI 9229/9239 are secure.

(NI 9229/9239 with screw-terminal)

 Use shielded copper wire for all cable connections 

to the device.

Use twisted-pairs wire to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ± 5 °C.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the NI 9229/9239 measurement 
circuitry is at a stable operating temperature.

Initial Setup

Complete the following steps to set up the NI 9229/9239.
1.

Install the NI-DAQmx driver software.

2.

Make sure the NI cDAQ-9178 power source is not connected.

Table 1.  

Recommended Equipment

Equipment

Recommended 

Model

Parameter 

Measured

Requirements

Calibrator

Fluke 5700A

Accuracy,

Adjustment

Use a high-precision voltage source 
with an accuracy 

100 ppm and an 

output impedance 

50 

.

Function 

Generator

NI 4461

Gain Matching,

Phase Matching

Use a function generator capable of 
sourcing both 1 kHz and 20 kHz 
±10 V

pk

 AC sinusoidal signals with 

amplitude uncertainty of ±10%.

Chassis

NI cDAQ-9178

All

Summary of Contents for NI 9229

Page 1: ...y Verification 4 Gain Matching Verification 6 Phase Matching Verification 7 Adjustment 9 Accuracy Adjustment 9 EEPROM Update 10 Reverification 10 Accuracy Under Calibration Conditions 10 Worldwide Sup...

Page 2: ...nstructions and Specifications NI 9229 9239 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabV...

Page 3: ...ctions to the device Use twisted pairs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C Keep relative humidity below 80 Allow a warm up time of at least 10 minut...

Page 4: ...elf Test to ensure that the device is working properly Verification The following performance verification procedures describe the sequence of operation and provide test points required to verify the...

Page 5: ...Accuracy Verification Module Physical Channels Min V Max V Scaled Units Terminal Configuration 9229 AI0 3 60 60 Volts Differential 9239 AI0 3 10 10 Volts Differential Table 3 NI 9229 9239 Timing Conf...

Page 6: ...ure function generator according to Table 6 3 Acquire and average samples a Create and configure an AI voltage task on the NI 9229 9239 according to Table 7 Table 6 Function Generator Configuration fo...

Page 7: ...If the Gain Matching verification procedure determines that the NI 9229 9239 is outside of the limits refer to Worldwide Support and Services for assistance in returning the device to NI Phase Matchin...

Page 8: ...able the function generator g Clear the task 5 Calculate the phase for each channel on the NI 9229 9239 NI recommends using the Extract Single Tone Information VI to calculate phase Table 9 Function G...

Page 9: ...r as shown in Figure 1 Note If the calibrator outputs are truly floating the negative output must be connected to a quiet earth ground as well as AI to give the entire system a ground reference 3 Adju...

Page 10: ...determine the As Left status of the device Note If any test fails Reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to...