10. Close the door slowly.
11. Tighten the four captive screws on the door of the device to 1.3 N · m (11.5 lb · in.).
Using the Reset Button on the MON-10467 (DC)
Caution
Remove power to the MON-10467 (DC) before opening it for service.
Attention
Mettez le MON-10467 (DC) hors tension avant de l'ouvrir pour
l'entretien.
To restart the device, cycle power to the device.
To delete the stored password and gateway bonding information from the device, complete the
following steps.
1.
Turn off external power to and completely remove power from the device.
2.
Loosen the four captive screws on the door of the device with a 1/4 in. socket or 6 mm
(1/4 in.) flat-blade screwdriver.
3.
Open the door of the device slowly.
4.
Press and hold the reset button.
5.
Apply DC power while holding the reset button.
6.
Release the reset button.
7.
Clean the gasket and gasket mating surface with a dry or damp cloth.
8.
Close the door slowly.
9.
Tighten the four captive screws on the door of the device to 1.3 N · m (11.5 lb · in.).
Using the Reset Button on the MON-10467 (BATTERY)
To restart the device, complete the following steps.
1.
Loosen the four captive screws on the door of the device with a 1/4 in. socket or 6 mm
(1/4 in.) flat-blade screwdriver.
2.
Open the door of the device slowly.
3.
Press the reset button for one second while the device is powered on.
4.
Clean the gasket and gasket mating surface with a dry or damp cloth.
5.
Close the door slowly.
6.
Tighten the four captive screws on the door of the device to 1.3 N · m (11.5 lb · in.).
To delete the stored password and gateway bonding information from the device, complete the
following steps.
1.
Loosen the four captive screws on the door of the device with a 1/4 in. socket or 6 mm
(1/4 in.) flat-blade screwdriver.
2.
Open the door of the device slowly.
3.
Disconnect the battery pack cable from the battery power connector port.
4.
Press and hold the reset button.
5.
Insert the battery pack cable into the battery pack power connector port.