background image

NI 9205/9206 Calibration Procedure

6

ni.com

4.

Adjustment Procedure

—Perform an external calibration that adjusts 

the device calibration constants with respect to a known voltage 
source.

5.

Perform another verification to ensure that the device operates within 
its specifications after adjustment.

The first four steps are explained in the following sections.

Initial Setup

You must configure the device in Measurement & Automation Explorer 
(MAX) to communicate with NI-DAQmx.

Complete the following steps to configure a device in MAX:

1.

Install the NI-DAQmx driver software.

2.

Make sure that no power is connected to the module terminals. If the 
system is in a nonhazardous location, the chassis power can be on 
when you install the module.

3.

Insert the module into an available slot in the cDAQ-9172 chassis.

4.

Launch MAX.

5.

Right-click the device name and select 

Self-Test 

to ensure that the 

device is working properly.

Note

When a device is configured with MAX, it is assigned a device name. Each function 

call uses this device name to determine which DAQ device to calibrate. This document uses 

dev1

 to refer to the device name. In the following procedures, use the device name as it 

appears in MAX.

Self-Calibration

Self-calibrate the device before completing the first verification, after the 
device has warmed up for the recommended time period. 

DAQmxSelfCal

 

(DAQmx Self Calibrate VI), as shown in the following table, measures the 
onboard reference voltage of the device and adjusts the self-calibration 
constants to account for any errors caused by short-term fluctuations in the 
operating environment. When you self-calibrate a device, no external 
signal connections are necessary.

NI-DAQmx Function Call

LabVIEW Block Diagram

Call 

DAQmxSelfCal

 with the 

following parameters:

deviceName

dev1

Summary of Contents for 9205

Page 1: ... interval as defined by the measurement accuracy requirements of your application National Instruments recommends that you routinely perform a complete calibration at least once every two years You can shorten this interval based on the accuracy demands of your application or requirements of your processes Self calibration can be performed as needed or when the temperature varies by 5 C or more fr...

Page 2: ... to important information bold Bold text denotes items that you must select or click in the software such as menu items and dialog box options Bold text also denotes parameter names and hardware labels italic Italic text denotes variables emphasis a cross reference or an introduction to a key concept Italic text also denotes text that is a placeholder for a word or value that you must supply monos...

Page 3: ...hanges all EEPROM constants it invalidates the original calibration certificate If an external calibration is done with a traceable signal generator source a new calibration certificate can be issued This document describes specific programming steps for writing an external calibration procedure for the NI 9205 9206 Software and Documentation Requirements This section describes the software and do...

Page 4: ...t measurement concepts To access this help file select Start All Programs National Instruments NI DAQ NI DAQmx C Reference Help DAQ Getting Started Guide for NI DAQ 8 0 or later This guide describes how to install NI DAQmx for Windows software and NI DAQmx supported DAQ devices and how to confirm that your device is operating properly To access this guide select Start All Programs National Instrum...

Page 5: ...ire for all cable connections to the device Use twisted pair wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 10 C The device temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the measurement circuitry is at a stable operating temperature Calibration Process This secti...

Page 6: ...72 chassis 4 Launch MAX 5 Right click the device name and select Self Test to ensure that the device is working properly Note When a device is configured with MAX it is assigned a device name Each function call uses this device name to determine which DAQ device to calibrate This document uses dev1 to refer to the device name In the following procedures use the device name as it appears in MAX Sel...

Page 7: ...fy all values although you can save time by verifying only the values used in your application 3 If you use C function calls create a task using DAQmxCreateTask as shown in the following table If you use LabVIEW skip this step The task is created in step 4 in LabVIEW Table 1 NI Connection Accessories Device Connection Accessory Connection Instructions NI 9205 9206 NI 9940 36 position strain relief...

Page 8: ...x Timing VI as shown in the following table NI DAQmx Function Call LabVIEW Block Diagram Call DAQmxCreateAIVoltageChan with the following parameters taskHandle taskHandle physicalChannel dev1 ai0 nameToAssignToChannel myVoltageChannel terminalConfig DAQmx_Val_Cfg_Default minVal 10 0 maxVal 10 0 units DAQmx_Val_Volts customScaleName NULL NI DAQmx Function Call LabVIEW Block Diagram Call DAQmxCfgSam...

Page 9: ... Task VI as shown in the following table 10 Repeat steps 3 through 9 for all ranges and all values 11 Disconnect the calibrator from the device or proceed to the Adjustment Procedure section to adjust the analog input calibration constants NI DAQmx Function Call LabVIEW Block Diagram CallDAQmxStartTask with the following parameter taskHandle taskHandle NI DAQmx Function Call LabVIEW Block Diagram ...

Page 10: ...s adjusted by the metrology laboratory The default password is NI Complete the following steps to adjust the device with a calibrator 1 Refer to Table 1 for a list of the NI connection accessories for the NI 9205 9206 and instructions on connecting the calibrator to the device 2 Set the calibrator to output 4 0 V 3 Open a calibration session on the device using DAQmxInitExtCal DAQmx Initialize Ext...

Page 11: ...ard memory Note If the device fails the calibration procedure no constants are written to the EEPROM 6 Disconnect the calibrator from the device The device is now calibrated with respect to the external source After calibrating the device verify the analog input operation by repeating the Verification Procedure section NI DAQmx Function Call LabVIEW Block Diagram Call DAQmxMSeriesCalAdjust with th...

Page 12: ...0 4 0 2 114 10 4 0 1 Absolute accuracy values at full scale on the analog input channels assume the device is operating within 10 C of the last external calibration and are valid for averaging 100 samples immediately following internal calibration 2 Sensitivity is the smallest voltage change that can be detected It is a function of noise Table 3 Accuracy Details Nominal Range V Residual Gain Error...

Page 13: ...ResidualAIGainError GainTempco TempChangeFromLastInternalCal ReferenceTempco TempChangeFromLastExternalCal OffsetError ResidualAIOffsetError OffsetTempco TempChangeFromLastInternalCal INL_Error Noise Uncertainty RandomNoise 3 for a coverage factor of 3 Σ and averaging 10000 points ResidualAIGainError ErrorRelativetoExternalRefppm ReferenceDriftOverTimeppm_per_year GainTempco AnalogCircuitryGainTem...

Page 14: ... when the device is within its 2 year calibration interval the test point value should fall between these upper and lower limit values Table 4 NI 9205 9206 Analog Input Values Range V Test Point 2 Year Limits Minimum Maximum Location Value V Lower Limit V Upper Limit V 10 10 Max 9 980000 9 976836 9 983164 10 10 Mid 0 000000 0 001407 0 001407 10 10 Min 9 980000 9 983164 9 976836 5 5 Max 4 990000 4 ...

Page 15: ...ruments also has offices located around the world to help address your support needs For telephone support in the United States create your service request at ni com support and follow the calling instructions or dial 512 795 8248 For telephone support outside the United States contact your local branch office Australia 1800 300 800 Austria 43 0 662 45 79 90 0 Belgium 32 0 2 757 00 20 Brazil 55 11...

Reviews: