National Instruments Corporation
11
NI 9205/9206 Calibration Procedure
4.
Perform an external adjustment using
DAQmxMSeriesCalAdjust
(DAQmx Adjust M-Series Calibration VI), as shown in the following
table.
5.
Save the adjustment to the EEPROM using
DAQmxCloseExtCal
(DAQmx Close External Calibration), as shown in the following table.
This function also saves the date, time, and temperature of the
adjustment to the onboard memory.
Note
If the device fails the calibration procedure, no constants are written to the
EEPROM.
6.
Disconnect the calibrator from the device.
The device is now calibrated with respect to the external source. After
calibrating the device, verify the analog input operation by repeating the
Verification Procedure
section.
NI-DAQmx Function Call
LabVIEW Block Diagram
Call
DAQmxMSeriesCalAdjust
with the following
parameters:
calHandle
:
calHandle
referenceVoltage
:
4.0
NI-DAQmx Function Call
LabVIEW Block Diagram
Call
DAQmxCloseExtCal
with the following
parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_
Action_Commit