Adjustment Procedure
The NI 9205/9206 adjustment procedure adjusts the analog input
calibration constants. At the end of each adjustment procedure, the new
constants are stored in the external calibration area of the device EEPROM.
These values are password-protected, which ensures that you do not
accidentally access or modify any calibration constants adjusted by the
metrology laboratory. The default password is
NI
.
Complete the following steps to adjust the device with a calibrator:
1.
Refer to Table 1 for a list of the NI connection accessories for the
NI 9205/9206 and instructions on connecting the calibrator to the
device.
2.
Set the calibrator to output 4.0 V.
3.
Open a calibration session on the device using
DAQmxInitExtCal
(DAQmx Initialize External Calibration VI), as shown in the following
table.
NI-DAQmx Function Call
LabVIEW Block Diagram
Call
DAQmxInitExtCal
with the following
parameters:
deviceName
:
dev1
password
:
NI
calHandle
:
&calHandle