Rev 8, Dated 27-2-2018
38
Fig 1
1. When the treadmill reaches to its maximum elevation press ‘
OK
’
button. When ‘
OK
’ button is pressed treadmill starts moving down and
following message as shown in fig 2 is displayed.
Fig 2
2. After the treadmill reaches its minimum elevation press ‘
OK
’ button.
When ‘
OK
’ is pressed calibration constants for grade are calibrated and
stored. These calibration constants are displayed in ST-Win configuration
dialog box.
3. If an error is detected this message appear on the screen,
Fig 3
B)
Calibrate speed
For calibrating the speed of the treadmill go to ‘
Utility
’ menu and select
‘
Calibrate speed
’ option. This will display the dialog box as shown in the
figure below. Count the number of seconds for 14 rotations and enter the
number in the edit box. Do this for one more time and enter the count in
the second edit box. Press ‘
OK
’ button to move to next stage of the
protocol.