12 Spatial Averaging
150
SRM-3006
Narda
To pause the measurement:
6.
Press the
Pause
softkey.
ª
The instrument stops averaging (No. of SAVG is static) and displays
the averaged result. It continues measuring in the background
(indicated by No. of Runs) but does not use these values for
determining the average.
7.
If required, move to a new position in the room and press the
Continue
softkey.
ª
The instrument resumes averaging (indicated by No. of SAVG).
8.
Move the antenna along the next path to be measured in the room.
9.
You can repeat steps 6 through 8 as often as necessary until you have
the final result.
To save the result in the measurement data memory:
⇒
Press the
SAVE
key.
To delete the result:
⇒
Press the
Clear
softkey.
To exit from Spatial Averaging:
⇒
Press the
ESC
key.
Note:
The instrument saves the actual (current) result of spatial averaging
in the measurement data memory when you press the
SAVE
key.
You can therefore use this function to save intermediate results.
When you press the
Continue
softkey, the instrument resumes
averaging and takes account of all the measured values obtained so
far. The instrument does not clear all the measurement values
recorded and start a new spatial averaging procedure until you press
the
Clear
softkey.
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......