17 Specifications
184
SRM-3006
Narda
17.1.2
Modes
Spectrum Analysis mode
Measurement principle Spectrum analysis
Resolution bandwidths
(RBW) (-3 dB)
10 Hz to 20 MHz (in steps of 1, 2, 3, 5, 10, 20)
List of available RBWs depends on selected sweep SPAN
Measurement range
setting (MR)
• Set individually from a list
or
• using the
MR Search
function for determining the optimum
measurement range at a given time
Video bandwidth
(VBW)
0.2 Hz to 2 MHz (depending on the selected RBW)
Filter
• Type: Gaussian
• Shape factor (- 3 dB/- 60 dB):
< 3.8 (for RBW
≤
100 kHz)
Evaluation functions
• ACT: Displays current (actual) spectrum
• MAX: Maximum hold function
• AVG: Average over a selectable number of spectra (4 to 256) or a
selectable time period (1 to 30 minutes)
• Max AVG: Maximum hold function after averaging over a defined
number of spectra
• Min: Minimum hold function
• Min AVG: Minimum hold function after averaging over a defined number
of spectra
• Standard: Display of the selected safety standard.
Marker functions
• Delta marker on one Result Type or for displaying the difference
between two Result Types
• Highest peak, next peak right, next peak left, next higher peak, next
lower peak
• Marker field (frequency, level, service name according to the selected
service table)
Evaluation functions
• Peak table (list of 50 highest peaks)
• Integration over a user-specified frequency range
Axis
• Isotropic measurement (isotropic result displayed directly)
• Measurement of X-, Y- or Z- axis (separate measurement of a single
axis using the isotropic / three axis antenna)
Zoom
• Zoom Min: Sets the lower frequency limit of the zoom window
• Zoom Max: Sets the upper frequency limit of the zoom window
• Zoom Cent: Moves the zoom window along the frequency axis
• Zoom Span: Changes the scale of the zoom window
• Execute Zoom: Sets the zoom window limits to the selected frequency
values
Display functions
• Y-scale range 20, 40, 60, 80, 100 or 120 dB
• Y-scale reference MR -100 dB to MR + 20 dB (-130 dB to 40 dBm)
• Screen arrangement: Enlarges result display area by hiding other
information.
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......