10 Scope Mode
Narda
SRM-3006
131
10.4
Using the trigger function
The trigger function allows you to synchronize the measurement cycle with
the signal. The result is that after triggering, the display appears “frozen” so
that the signal characteristics can be seen and you can measure the
individual signal components, for example.
Note:
You cannot generate a screenshot if the trigger function has been
selected but not started. You must either start the trigger (
Arm
softkey) or exit from the Trigger menu to generate a screenshot.
You can set the following conditions:
Setting the trigger function:
1.
Press the
Trigger Mode
softkey and select the appropriate function.
2.
Press the
Trigger Level
softkey and use the rotary control to set a
threshold value.
ª
The threshold is indicated by a horizontal blue line in the display.
3.
Press the
Trigger Edge
softkey to select the edge (toggle function):
– Softkey is labeled
Falling
: rising edge is selected (= Rising).
– Softkey is labeled
Rising
: falling edge is selected (= Falling).
4.
Press the
Trigger Delay
softkey and select a value:
– Negative value: earlier time
– Positive value: later time
ª
When you select a negative value, the trigger delay will be shown as
a vertical blue line in the display.
Table 36: Trigger conditions
Condition
Explanation
Trigger Mode
Trigger function:
• Free Run: no trigger function
• Single: The display is frozen after the first valid event.
• Multiple: The display is frozen again after each new valid
event.
Trigger Level
Level activating the trigger
Trigger Delay
Delay in activating the trigger
Trigger Edge
Signal edge activating the trigger:
• Rising
• Falling
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......