4-4 Applications
4.1.3 Test Setup
Fig. 7-2. shows an example of test setup for RFI voltage measurement.
The DUT is placed 0,4 m from an horizontal or vertical earthed conducting
surface of at least 2 m x 2 m in size.
A table top DUT is placed 0,8 m from the LISN and at least 0,8 m from any
other earthed conducting surface. If the measurements are made in a
shielded room, the DUT shall be placed 0,4 m from one of the walls of the
room.
The LISN shall be bonded to the reference conducting surface.
A Floor standing DUT is placed 0,1 m above an horizontal earthed
conducting surface of at least 2 m x 2 m in size. This size shall be
exceeding by at least 0,5 m the projection of the DUT on the conducting
surface. The power cable should be 1m long; longer cable should be
centrally bundled for at least 40 cm.
DUTs without a PE (Protective Earth) conductor and manually operated
DUTs shall be measured in conjunction with an auxiliary screen or an
“Artificial Hand”, as duly specified in the relevant standards.
All the details and information on the test setup are written on the latest
version of the applicable Standard.
4.1.4 Guidance on a
preliminary
Measuring
Procedure
A step-by-step example of a conducted test manually performed is the
following:
1. Switch ON the PMM 7010 and enter in
Sweep
Mode pressing the
relevant button;
2. Select
Display
and then
120dB
; Reference Level should be between
110 and 135 dB
μ
V; press
Esc
;
3. Touch the
Measure
key, then
Freq
B Band
;
4. Then
Level
,
Input
and set the Minimum Attenuation to 10 dB and the
Attenuation is automatic; press
Esc
;
5. The
Pulse Limiter
is always enabled to protect the input from voltage
spikes;
6. Enter in the
Detector
menu and select
Peak
; press
Esc
two times;
7. Connect the internal LISN or any other suitable transducer to the RF
input, selecting the proper phase/line whenever applicable;
8. With the EUT switched OFF touch
Exec Sweep
and wait until the first
scan has been finished;
9. Carry out an overview of the whole spectrum and check the by
executing a pre-measurement with peak detector that the ambient RF
disturbances are at least 20 dB below the desired emission levels;
10.
If the ambient noises are low enough continue to the next step,
otherwise take all the necessary steps to reduce the ambient
disturbances (e.g. go to a shielded room, etc.);
11. Switch ON the EUT and then
Measure
,
Exec Sweep
again and wait
until the scan has been finished;
12. Enter in the
Display
menu and adjust the Dynamic Range and the
Reference Level to the most convenient visualization;
13. If the results are satisfactory select
Marker
,
Peak
and then either
Analyzer
or
Tune
to enter respectively in Spectrum Mode or in Manual
Mode and deeply investigate the highest signals;
14.
If the results of the scan need to be refined, select a
Limit
, if
appropriate, and then
Measure
,
Level
and
Smart Detector
(if a limit
has been loaded);
15. In the
Smart Detector
menu check the best combination among those
listed; enter a
Margin
, if needed; press
Esc
;
16.
Exec Sweep
again and repeat from step #12.
Summary of Contents for PMM 7010 Series
Page 8: ...VIII Contents This page has been left blank intentionally...
Page 36: ...4 6 Applications This page has been left blank intentionally...
Page 50: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Page 76: ...Suggerimenti Commenti Note Suggestions Comments Note...