4
STEP 4
At the end of set time interval an alarm will
sound. The alarm will continue to sound until
the power switch is set to "OFF" position. The
test chamber will automatically vent and stop
heating. Allow the chamber temperature to
decrease to at least 100°C (212°F) before
opening the door. Open the door and leave ajar.
STEP 5
Before the next cycle "drain condensate".
Condensate should be drained after each cycle.
The chamber reservoir should be refilled to
water level indicator.
PROCEDURE FOR OPERATION OF
DRYER CYCLE
To operate the dryer cycle: leave door slightly
ajar, set the dryer switch to "ON" position, set
the dryer temperature control to "LOW" position,
set the timer to desired drying time interval, set
the power switch to "ON" position. At the end of
the set time interval, an alarm will sound. The
alarm will continue to sound until the power
switch is set to "OFF" position. The dryer
temperature control also has a medium and high
setting, select a higher temperature if desired
but use caution to avoid overheating contents.
MAINTENANCE SUGGESTIONS
1) At the end of one or two test chamber
cycles: Remove screw fastener which holds
shelf in place. Remove shelf to gain access
to water reservoir. Add mild detergent to
water and scrub reservoir and rinse cham-
ber thoroughly with clear water. Replace
shelf as above.
2)
Oil "T" handle threads twice a year or as
often as needed.